Re: [AMMRL] BACS-120 pincher

From: Perry Pellechia <pellechia_at_sc.edu>
Date: Mon, 4 Dec 2023 14:40:48 +0000

Justin,
This is an easy fix. There is an o-ring under the back cylinder that is broken
and needs to be replaced. Remove the three screws on the side and carefully
slide the black cylinder off. The 5 fingers are free to fall off their pivots
since the broken o-ring is not holding them to the arm anymore, so you have
to be careful they do not fall (into the magnet). The new o-ring slides over
the finger and hold them open and in place to the arm.

Perry.
--

Perry J. Pellechia
Director of NMR Services
Research Professor
University of South Carolina
Department of Chemistry and Biochemistry




> On Dec 4, 2023, at 8:38 AM, Justin Pontius via groups.io <jpontius=nd.edu_at_groups.io> wrote:
>
> Hello,
>
> We have an aging Avance2 400 with a BACS-120 sample changer. Recently the
> pincher seems to get fouled up when trying to pick up tubes. When the "fingers"
> open they dangle freely. When it moves over the tube every now and
> then a finger will get pushed to it closed position. Of course the changer
> detects an error and stops. I'm wondering if this is normal or did something
> break inside the mechanism. I'm terrified to try and take it apart. It's
> mostly working now but the fear is if I get it apart then it won't ever go
> back together again. Does anyone have any experience with it? I'll attach
> some pics and a video below.
>
> https://photos.app.goo.gl/QxJHZHZyJDnvLABu6
>
> --
> Justin Pontius
> NMR Facility Engineer
> Magnetic Resonance Research Center
> University of Notre Dame
> nmr.nd.edu
> 574-631-7601 (direct line)
>
> 145A Nieuwland Science Hall
> Notre Dame, IN 46556
>
>



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#809): https://ammrl.groups.io/g/main/message/809
Mute This Topic: https://groups.io/mt/102970731/7559972
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Dec 04 2023 - 06:57:12 MST

This archive was generated by hypermail 2.4.0 : Mon Jan 01 2024 - 19:06:37 MST