While not the exact situation, we upgraded two 400 MHz wide bore Oxford magnets
from a Varian (Infinity Plus) to Bruker NEO. We use both Bruker and Phoenix
narrow bore probes (2.5, 3.2, and 4mm) without much issue. We also upgraded a
Varian Inova 600 MHz to a Bruker NEO. That has a 1996 Oxford unshielded narrow
bore. We use 1.3, 3.2, and 4mm probes. The only thing I’ve found is
that inserting and ejecting is more challenging on these old magnets. They are
not a deal breaker, just requires a little tap to help out. We have a new 800
magnet that does not suffer the same issues.
Dan
Michigan State University
-=-=-=-=-=-=-=-=-=-=-=-