Re: [AMMRL] Topspin acquisition license expired (?)

From: Shin, Ronald via groups.io <shinr=uab.edu_at_groups.io>
Date: Tue, 18 Aug 2026 13:20:10 +0000

William,

Your Avance III console should allow you to upgrade your Topspin installation
to 3.7.

I believe that it was either 3.5 or 3.6, that replaced the FlexLM licensing
system with the Code Meter licensing system. If your original license file
is still valid, then you should be able to upgrade to the newer version of
Topspin, and that may "reset" the expiration date with the new licensing
system.

Ronald Shin
University of Alabama at Birmingham
Central Alabama High Field NMR Facility


________________________________
> From: main_at_ammrl.groups.io on behalf of William Jarrett via groups.io
> Sent: Monday, August 17, 2026 3:40 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Topspin acquisition license expired (?)

Colleagues:

   This concerns an Avance III 600 install in 2012 and using Topspin
3.1pl7. I was help a user set up a VT NMR acquistion when I got the
following error message:

          No license for TOPSPIN acquisition

Examination of the license file indeed indicate that although every
other module was dated 13-jun-2028, the acquitision was dated
29-jul-2026. This is especially frustrating as the expiration warning
issued when starting Topspin only stated the 2028 timelimit. Does
anyone know if there is a simple fix, or am I in for a long semester?

Regards,
WLJ

--
Dr. W. L. Jarrett
University of Southern Mississippi
School of Polymer Science and Engineering
Phone: (601)266-4551


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3696): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3696__;!!PvDODwlR4mBZyAb0!SRzR-q-5_DFcVy2M7Os89IIYloRdvJ=
gcoa9jHLmP5mGITzyD0ThR9tp0zZTSLLI2L8umoZ8J-Ty5vnsfz5TJUqZX$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120806311=
/7559972__;!!PvDODwlR4mBZyAb0!SRzR-q-5_DFcVy2M7Os89IIYloRdvJgcoa9jHLmP5mGIT=
zyD0ThR9tp0zZTSLLI2L8umoZ8J-Ty5vnsfz6Gkefa-$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Aug 18 2026 - 06:59:05 MST

This archive was generated by hypermail 2.4.0 : Tue Sep 01 2026 - 13:53:27 MST