[AMMRL] Bauer Vertical High Pressure Gas Compressor Units G 120 II problem

From: Wang, Tianzhi via groups.io <tiawang=utmb.edu_at_groups.io>
Date: Fri, 14 Aug 2026 16:04:09 +0000

Dear All

My vertical High Pressure BAUER Gas Compressor, which operates within a helium
recovery system, has an intermittent low inlet pressure fault. Our local
authorized distributor has evaluated the system but has been unable to resolve
the issue.

Problem Description:

During normal operation, the compressor runs with a steady inlet pressure of
-0.2 PSI. Intermittently—at intervals ranging anywhere from 10 minutes to
100 hours of runtime—the compressor suddenly trips due to a low inlet
pressure fault.

Diagnostic Details:

  * Pressure Readings: When the system trips, the digital controller display
indicates that the pressure has suddenly dropped to -18.8 PSI. However,
the mechanical pressure gauge on the inlet line remains completely stable at
-0.2 PSI.
  * System Reset: Powering off the system completely and restarting it
temporarily clears the fault, allowing the compressor to run again for an
unpredictable duration before the fault repeats.
  * Preliminary Troubleshooting: We previously identified and secured a loose
24V emergency stop wire, which briefly seemed related, but the issue has returned.
We suspect that there may be an intermittent connection failure in the 24V power
wiring or that the pressure sensor itself is failing under operational vibration.

Given that the mechanical gauge confirms the system is not actually pulling
a deep vacuum, this appears to be an electrical or signal loop anomaly.
I appreciated any suggestions

Best,
Tainzhi
-------------------------------------------------------------------
Tianzhi Wang, Ph. D.
NMR Manager
Sealy center for Structural Biology & Molecular Biophysics
University of Texas Medical Branch,
301 University Bvld, NMR dockside building, Rom 1.104
Galveston, TX 77555-1157
Email: tiawang_at_utmb.edu
(409)7476821 (O)



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3697): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3697__;!!PvDODwlR4mBZyAb0!XxBoUpDm9-lcaAwshvrgKw2UHovagN=
HGPEaLZQK3_6rrd_7o3_lt0e9z9MfQ_5PpiCcW2g8WQHV3rZ0OGV5yXqJePdT5$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120808287=
/7559972__;!!PvDODwlR4mBZyAb0!XxBoUpDm9-lcaAwshvrgKw2UHovagNHGPEaLZQK3_6rrd=
_7o3_lt0e9z9MfQ_5PpiCcW2g8WQHV3rZ0OGV5yXiGd-dMe$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Aug 18 2026 - 07:09:04 MST

This archive was generated by hypermail 2.4.0 : Tue Sep 01 2026 - 13:53:27 MST