Am 14.08.2026 um 21:48 schrieb Uddigiri, Venkateswara Reddy via groups.io <venkateswarareddy.uddigiri=louisville.edu_at_groups.io>:
Hi Bert,Thank you for your help.Regarding your suggestions:• Check whether all shims are set (not only Z shims) and post a screenshot of the "Shims" panel.
I have uploaded the shim profile screenshot.• The shim cable may not be correctly connected to the shim power supply, or a fuse at the back of the power supply may have blown. You did not specify the console type.
The system is a DD2, and I have uploaded an image of the console and setup.• Please insert the D₂O or Autotest sample, perform an autolock, then go to "Tools → Standard Calibration Exp. â†= ’ Setup Gradient Shimming" and record the trial profiles. Type "ff dss" in the command line and post a screenshot.
I have attached the requested screenshots.• The NMR probe may have moved (typically downward) relative to the shim tube, which would be visible as a misalignment of the two PFG profiles.
The probe was recently repaired, and unfortunately the previous manager did not mark its original position. I have tried several insertion depths, but I still observe the same broad underlying pattern. Images collected at different probe positions are attached.• If you see dips in the profiles (especially the lower profile), there may be dirt inside the probe. In that case, remove the probe and clean the sample bore with cotton swabs and isopropyl alcohol.
Do you think it would be worthwhile to test the probe while empty to determine whether any signal might be arising from contamination or dirt inside the probe? I plan to clean the probe next and will report back with the resu= lts.Thank you again for your assistance.Best regards,
Venky