Re: [AMMRL] Fix for daily ICON-NMR hangs / "too many files" error (Topspin 5.0 SP4)

From: Akien, Geoffrey via groups.io <g.akien=lancaster.ac.uk_at_groups.io>
Date: Mon, 10 Aug 2026 13:42:20 +0000

Hi folks,

If you are dealing with those IconNMR problems, then this may well be the
right course of action.

Having said that, in my hands service pack 4 on Windows 11 has broken WaveMaker's
detection of which channel pulses are supposed to go on, specifically only with
the new-style syntax.

If you're using e.g. selgpse, then an attempt to use wvm with WAVEMAKER0
create_shape(sp2, rsnob, length=40 ms)
...will fail with "Missing channel specification for shape defined outside
of pulse program". The same thing will happen with any pulse sequence containing
the same WaveMaker definition.

One workaround is to explicitly specify the channel and pulse e.g.
create_shape(sp2, rsnob, length=40ms, channel=f1, pulse=p12), but this affects
rather a lot of recent pulse sequences in the library if you're using those.
Another alternative is to switch over to the old-style syntax e.g.
USERA1 sp2:wvm: rsnob(40 ms), but then tolerate the popups reminding that this
syntax is deprecated and you shouldn't be using it anymore.


Thanks to Ken for pointing out where to look, and showing that the service
pack installer retains all original versions of files it has replaced, and
appends them with ".backup." In this case, restoring the old version of wvm.exe.backup
in <TSHOME>\prog\mod seems to fix things. I assume this was changed for a reason,
but a WaveMaker bug was not in the release letter so no idea what you give up
by restoring the original file.

Thanks

Geoff

> From: main_at_ammrl.groups.io On Behalf Of Kenneth Sharp-Knott via groups.io
> Sent: 10 August 2026 14:05
> To: main_at_ammrl.groups.io
> Subject: [External] [AMMRL] Fix for daily ICON-NMR hangs / "too many files" error (Topspin 5.0 SP4)


This email originated outside the University. Check before clicking links o=
r attachments.

Dear AMMRL colleagues,

If you have been dealing with the long-standing bug in ICON-NMR that requires
daily restarts of the automation queue, I highly recommend installing Service
Pack 4 for Topspin 5.0.

Historically, this bug caused random, silent hangs in older versions of Topspin.
In Topspin 5.0, it began manifesting as a "too many files" error. Since updating
to SP4, my ICON crashes and hangs have been reduced by at least 95%.

For those curious or running older versions, the patch seems to center on a
modification to this specific file:

/opt/Bruker/topspin5.0.0/prog/tcl/libtix/iconnmr/bruklib/Ez_Lib.tcl

It is highly likely that this is a legacy bug. If that Tcl script exists in
your older Topspin installations, it might be possible to apply the fix
backward to earlier versions.

Best,

Ken Sharp-Knott
Viriginia Tech Dept of Chemistry
NMR Facility Manager



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3677): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3677__;!!PvDODwlR4mBZyAb0!RHybO9xfGu-X-Ss2c-1RUAi9nlOcM0=
yODzV5SqrSacP-ECroeVLZyYsFJPxoHMIjVTM9xiQR3_AR-KByMREbtFH4-Fuc$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120684220=
/7559972__;!!PvDODwlR4mBZyAb0!RHybO9xfGu-X-Ss2c-1RUAi9nlOcM0yODzV5SqrSacP-E=
CroeVLZyYsFJPxoHMIjVTM9xiQR3_AR-KByMREbtOPuadK-$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Aug 10 2026 - 06:44:04 MST

This archive was generated by hypermail 2.4.0 : Tue Sep 01 2026 - 13:53:27 MST