[AMMRL] Fix for daily ICON-NMR hangs / "too many files" error (Topspin 5.0 SP4)

From: Kenneth Sharp-Knott via groups.io <kknott=vt.edu_at_groups.io>
Date: Mon, 10 Aug 2026 13:04:54 +0000

Dear AMMRL colleagues,

If you have been dealing with the long-standing bug in ICON-NMR that requires
daily restarts of the automation queue, I highly recommend installing Service
Pack 4 for Topspin 5.0.

Historically, this bug caused random, silent hangs in older versions of Topspin.
In Topspin 5.0, it began manifesting as a "too many files" error. Since updating
to SP4, my ICON crashes and hangs have been reduced by at least 95%.

For those curious or running older versions, the patch seems to center on a
modification to this specific file:

/opt/Bruker/topspin5.0.0/prog/tcl/libtix/iconnmr/bruklib/Ez_Lib.tcl

It is highly likely that this is a legacy bug. If that Tcl script exists in
your older Topspin installations, it might be possible to apply the fix backward
to earlier versions.

Best,

Ken Sharp-Knott

Viriginia Tech Dept of Chemistry
NMR Facility Manager


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3676): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3676__;!!PvDODwlR4mBZyAb0!Q8FC0devDbjr7JpNYkiFxI64LV-wtd=
hfHTkhaTT0FpFrVXjvRJVgpUMMo-0aAYSjA7nzaiOB-40ZaLjtA0nR8BD6$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120684220=
/7559972__;!!PvDODwlR4mBZyAb0!Q8FC0devDbjr7JpNYkiFxI64LV-wtdhfHTkhaTT0FpFrV=
XjvRJVgpUMMo-0aAYSjA7nzaiOB-40ZaLjtA6bduST4$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Aug 10 2026 - 06:09:33 MST

This archive was generated by hypermail 2.4.0 : Tue Sep 01 2026 - 13:53:27 MST