Re: [AMMRL] AMMRL: Intermittent SampleCasePlus (60) problem - stuck in state mid-to-in, mid-to-out, etc.

From: Eugenio Alvarado via groups.io <ealva=umich.edu_at_groups.io>
Date: Thu, 6 Aug 2026 09:25:02 -0400

Josh,

We also had a similar problem a couple of years ago. I removed all the
layers and completely removed the sticky lubricant from all of them with
isopropanol.  Then I applied a non-stick dry film lubricant from
Dupont.  We have no problems since then.

DuPont™ Non-Stick | DuPont Lubricants

Hope that helps,
Eugenio

Eugenio Alvarado, Ph.D.
NMR Facility Director
Chemistry Department, room 3500. University of Michigan
930 N. University Ave., Ann Arbor, MI 48109-1055.
Voice: (734)-764-9924


On 8/5/26 11:55 AM, Josh Kurutz via groups.io wrote:
> Hi AMMRL Community.
>
> I'd like your help addressing a recurring intermittent problem with
> our SampleCasePlus (60 samples). It occasionally times out as the
> carousel slides back and forth to change which of the three rings of
> samples will be under the transport tube. We get a “red bar” error, a
> failed sample, and the queue stops. The error message is along the lines:
> /“…Fatal Error…1000 STOP_CHANGER…Timeout while trying to move
> carousel. Move tray FSM is stuck in state 9: moving mid to in.”/
> Similar errors appear for moving between other ring positions (mid,
> in, out). Recovery by resetting the autosampler with the emergency
> stop button is immediate - there’s no apparent change in carousel
> position upon successful reset - like the carousel was just slow or
> needed another couple seconds to read its position.
>
> I’d like some help either fixing the recurring problem or managing m
> expectations. I’ve been focusing on the mechanical issues, but perhap
> there’s a setting in the HA web interface where I can increase the
> timeout time by a couple seconds.
>
> In the past, we’ve gained some relief by:
>
> *
> Cleaning the aluminum fins that cover the lower unit’s interior at
> the front and back margins. Last time we did this, we removed the
> fin assembly from the cover (but kept it intact) so we could place
> it flat on a clean bench top, and wiped down the fins repeatedly
> with 70%-isopropanol-laden paper towels. We took care to not
> disturb the side edges, which are apparently lubricated with a
> white grease.
> *
> Using dry kimwipes to wipe down the interior track along which the
> carousel (“tray”?) slides back and forth between its three
> positions. I then applied a light coat of Festo grease (the same
> as used to lubricate the slider rail atop the magnet) to that track.
> *
> Gently swabbing all sensors with cotton-tipped swabs briefly
> dipped in electronics-grade isopropanol, paying extra attention to
> the ones sensing mid/in/out positions.
>
>
> That cleaning helped a lot a few months ago, but over the last few
> weeks the problems have been occurring again, and incident frequency
> is increasing. I also noticed some white powder occasionally builds up
> on top of one part of a fin, right next to where the white carousel
> plastic protrudes above the aluminum. This powder doesn’t really look
> like a spilled sample - my suspicion is that something is rubbing or
> scraping the carousel plastic and leaving a residue, but I don’t have
> a good basis for thinking that, and I’d be hard-pressed to identify
> where that action might happen.
>
> The portion of the SampleCasePlus manual covering regular maintenance
> doesn’t appear say anything about maintaining the fins or lubricating
> the carousel’s sliding track inside the lower unit.
>
> I’d asked Bruker’s Rover support about this before, and I got some
> good insight. When I first got the units, the aluminum fins were
> lightly greased, so I thought I should occasionally wipe them down and
> re-grease them. An engineer corrected me on that, saying the grease
> doesn’t really have a function when the unit is working, so I should
> just remove the grease. However, I still figured the white grease at
> the sides looks important to reduce wear.
>
> Any ideas? Should I just start scheduling disassembly and cleaning
> every few months?
>
> Thanks. - Josh
>
>
> P.S. What does the “FSM” in the error message mean, with
> respect to the tray? I normally read this acronym with an entirely nontechnical
> meaning.
>
> --
> Josh Kurutz, PhD
> NMR Facility Manager, Chemistry Dept.
> University of Chicago
> https://chemnmrlab.uchicago.edu/
> jkurutz_at_uchicago.edu
>


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3667): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3667__;!!PvDODwlR4mBZyAb0!R5Jg6wtsTogm3y9tkEk015gSm6mW-k=
__fuRETnHeXusSgrarH5EHLCVRvtUe3S9q95UT4PuvhdjoHvyTy1INmDEzbE8$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120613914=
/7559972__;!!PvDODwlR4mBZyAb0!R5Jg6wtsTogm3y9tkEk015gSm6mW-k__fuRETnHeXusSg=
rarH5EHLCVRvtUe3S9q95UT4PuvhdjoHvyTy1INlCgOAOA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Aug 06 2026 - 06:52:01 MST

This archive was generated by hypermail 2.4.0 : Tue Sep 01 2026 - 13:53:27 MST