Re: [AMMRL] AMMRL: Intermittent SampleCasePlus (60) problem - stuck in state mid-to-in, mid-to-out, etc.

From: Paris Ning, PhD via groups.io <pning=isotope.com_at_groups.io>
Date: Wed, 5 Aug 2026 18:06:30 +0000

Hi Josh,

Adding to this – we had similar problems but the culprit was not
degrading polymer but dust/grim settled between each layer. This made the
movement from inner ring to outer ring slightly slower than what the software
was expecting. We asked the Bruker engineer to clean the case, which she did
by dissembling the cover etc., and all has been good so far.

Hope this helps.

Cheers,
Paris

> From: main_at_ammrl.groups.io On Behalf Of Gabriel Gonzàlez via groups.io
> Sent: Wednesday, August 5, 2026 1:43 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] AMMRL: Intermittent SampleCasePlus (60) problem - stuck in state mid-to-in, mid-to-out, etc.

Dear Josh,

We had some kind of error like that 2 or 3 years ago. The problem was that =
the metallic cover of the samplecase (the one with 60 holes for the positio=
ns) is made of two layers of metal with a polymer in between.

The polymer had degraded and became sticky impeding movement in an aleatory=
 way. We got a replacement cover, but the Bruker engineer advised us to lea=
ve it without cover.

Since we work without the cover, there has been no more problems.

Hope that helps. Good luck

Gabriel


Dr. Gabriel González Ferrer
Responsable RMN / NMR Manager
Institut Català d’Investigació Química (ICIQ) / Inst=
itute of Chemical Research of Catalonia (ICIQ)
Av. Països Catalans 16 – 43007 Tarragona (Spain)
Phone +34 977920207 / +34 686669420 – Fax +34 977920224
ggonzalez_at_iciq.es<mailto:ggonzalez_at_iciq.es> - https://urldefense.com/v3/__h=
ttp://www.iciq.es__;!!PvDODwlR4mBZyAb0!VL-kt_KH1q3ODFDTF_ySthTki9k4qAWwroFn=
T6dkpN6T3kQAqlTdUFI4xvVmx0pOmTpORJWysi1e-1hxVK0lH1GFzY1L$ <https://urldefen=
se.com/v3/__http://www.iciq.es/__;!!PvDODwlR4mBZyAb0!VL-kt_KH1q3ODFDTF_ySth=
Tki9k4qAWwroFnT6dkpN6T3kQAqlTdUFI4xvVmx0pOmTpORJWysi1e-1hxVK0lH4OyNFyX$ >


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3658): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3658__;!!PvDODwlR4mBZyAb0!VL-kt_KH1q3ODFDTF_ySthTki9k4qA=
WwroFnT6dkpN6T3kQAqlTdUFI4xvVmx0pOmTpORJWysi1e-1hxVK0lHyA0d6PF$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/120613914=
/7559972__;!!PvDODwlR4mBZyAb0!VL-kt_KH1q3ODFDTF_ySthTki9k4qAWwroFnT6dkpN6T3=
kQAqlTdUFI4xvVmx0pOmTpORJWysi1e-1hxVK0lH5hBK6dA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Aug 05 2026 - 11:11:39 MST

This archive was generated by hypermail 2.4.0 : Tue Sep 01 2026 - 13:53:27 MST