Re: [AMMRL] PTS Synthesizer issue on Agilent DD2 #agilent #Hardware #Troubleshooting

From: Spin-Doc <bert.heise_at_spin-doc.net>
Date: Tue, 19 Aug 2025 22:11:24 +0200

Dear David,

you didn't explicitly mention it but I guess you have a dual-channel PTS D620
synthesizer? Or is it a single-channel PTS620 plus a PTS320 for low-band?
In case you have a dual D620 you can try what Yong-Wah did - swap channels
1&2 (just swap BNC & RJ45 plugs and enter "su" - no need to swap the cards).
PTS have different internal synthesizers for different frequency ranges,
so it may well be that both high- and lowband work fine after a swap.
If you have only one single PTS620 then it is worthwhile replacing the two
large 10,000 uF sieve capacitors which may have dried up, inducing AC ripple
into the electronics and disturbing it sh**less.
Often PTS can be "repaired" like that. If that doesn't help you may need a
replacement PTS (let me know if you can't source one locally, I have these).

Best regards
Bert


> Am 19.08.2025 um 18:06 schrieb Jones, David (Pharm) via groups.io:
>
> Apparently my DD2 is different.
> Possibly an early DD2/vnmrs hybrid.
>
> We have external interface cards to the pts.
> David
>
> Associate Professor
> Dept. Of Pharmacology
> University of Colorado
> School of Medicine
> Tel (303)-724-3600
>
>> On Aug 19, 2025, at 8:50 AM, Jones, David (Pharm) via groups.io wrote:
>>
>>
>> If you have a spare PTS from an Inova it will work on the DD2. Just swap
>> the interface card. We swap them between instrument for testing. And we
>> have had several occasions when one of the channels in the PTS fails.
>>
>> If you don’t have the same model a 620 can be used. Just be aware
>> that some pTS were configured using the AUX channel to only deliver 520
>> or 620. So you may have use the main output and not the aux. it is rare
>> that they are full band on both channels. Dual channel 320s are common.
>>
>> You can put the “new” PTS on the top of the console if you
>> don’t have room and run the connectors outside to that.
>>
>> David
>>
>> Associate Professor
>> Dept. Of Pharmacology
>> University of Colorado
>> School of Medicine
>> Tel (303)-724-3600
>>
>>> On Aug 19, 2025, at 8:03 AM, Kim, Yong Wah via groups.io wrote:
>>>
>>> Hi Daniel,
>>>
>>> Sounds like a problem I had with my Inova 600MHz. I have a 4 RF channel
>>> and 2 X PTS620 synthesizers. So, I swapped the synthesizers and the problem
>>> is gone. The synthesizer output for H1 also improved dramatically. The
>>> H1 pulses shorten by 20% and the H1 tune pattern is back. Interestingly the o
>>> "defective" synthesizer works just fine at the lower frequencies. The synthesizer
>>> probably handles the high, H1 frequency and the low / broad band frequencies
>>> on different paths.
>>>
>>> Regards
>>>
>>> Yong-Wah Kim, PhD.
>>>
>>>
>


Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist

Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net

Representing:
Stelar Fast Field Cycling NMR Relaxometers (http://www.stelar.it)
MestreLab NMR Software (http://www.mestrelab.com)
Barthel HF Amplifiers (http://www.barthel-hf.de)


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2740): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2740__;!!PvDODwlR4mBZyAb0!W1i6AFKtCRuzndodKMvXpu7Ue4O-fY=
udQZ02giXStmig6xQ1wqgnd1leHZw2irmvbjQYY6Qmt-MQs0lcPljUsg$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114781283=
/7559972__;!!PvDODwlR4mBZyAb0!W1i6AFKtCRuzndodKMvXpu7Ue4O-fYudQZ02giXStmig6=
xQ1wqgnd1leHZw2irmvbjQYY6Qmt-MQs0mB6k0l6A$
Mute #troubleshooting:https://urldefense.com/v3/__https://ammrl.groups.io/g=
/main/mutehashtag/troubleshooting__;!!PvDODwlR4mBZyAb0!W1i6AFKtCRuzndodKMvX=
pu7Ue4O-fYudQZ02giXStmig6xQ1wqgnd1leHZw2irmvbjQYY6Qmt-MQs0lJqId_pg$
Mute #agilent:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/mu=
tehashtag/agilent__;!!PvDODwlR4mBZyAb0!W1i6AFKtCRuzndodKMvXpu7Ue4O-fYudQZ02=
giXStmig6xQ1wqgnd1leHZw2irmvbjQYY6Qmt-MQs0kZWWnKcg$
Mute #hardware:https://urldefense.com/v3/__https://ammrl.groups.io/g/main/m=
utehashtag/hardware__;!!PvDODwlR4mBZyAb0!W1i6AFKtCRuzndodKMvXpu7Ue4O-fYudQZ=
02giXStmig6xQ1wqgnd1leHZw2irmvbjQYY6Qmt-MQs0kKZ6qbCA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Wed Aug 20 2025 - 04:15:51 MST

This archive was generated by hypermail 2.4.0 : Mon Sep 01 2025 - 15:37:08 MST