Hi Luke and Kenneth,

In my setup (which looks exactly like Lukes) you can just simply take off
the inner aluminium ring and thus get access to the three screws holding
the bst. That way you do not need to touch the sample case  when
removing the bst. Just don't forget to put the ring back on (which ist
what I did ;) ). I did it together witha Bruker engineer here, so I am
pretty sure it is safe :)

Fingers crossed that it will work.

Best, André 

--

Dr. André Dallmann
Leiter der NMR-Abteilung
Email: andre.dallmann_at_chemie.hu-berlin.de
Raum: 0'103
Telefon: 82952
Handy: 01797657855O
RCID-ID: https://orcid.org/0000-0002-0141-2635
Webseite: https://www.chemie.hu-berlin.de/en/forschung-en/nmr-spectroscopy/nmr-spectroscopy?set_language=en

Adresse:
Humboldt Universität zu Berlin
Institut für Chemie
Leiter der NMR-Abteilung
Institut für Chemie
AG NMR
Brook-Taylor-Str. 212489
Berlin
Deutschland

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2711): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2711__;!!PvDODwlR4mBZyAb0!UXIEDgg91lBNASJK-LBykkXKQE6-OE=
XX31UKb9Ws5AjGzqWa7LV4AgfYRxOb6GAj_VThVYeFPqPPMU3Dnj1GGUMDekPl$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/114652860=
/7559972__;!!PvDODwlR4mBZyAb0!UXIEDgg91lBNASJK-LBykkXKQE6-OEXX31UKb9Ws5AjGz=
qWa7LV4AgfYRxOb6GAj_VThVYeFPqPPMU3Dnj1GGcp2Hx8B$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Aug 12 2025 - 04:56:19 MST

This archive was generated by hypermail 2.4.0 : Mon Sep 01 2025 - 15:37:08 MST