We designed a sturdy liquid nitrogen hose connected to a U-shaped stainless
steel component, featuring an adjustable tightening mechanism at the end.
This setup is both effective and foolproof, making it a worthwhile investment
for enhancing safety with a robust system.
http://nmr.vuw.ac.nz
School of Chemical and Physical Sciences
Victoria University of Wellington
New Zealand
________________________________
> From: main_at_ammrl.groups.io on behalf of jan.lang via groups.io
> Sent: Friday, 30 August 2024 02:49
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] Tubing for N2 fill exhaust?
Hi,
even though it „can NOT cope with the temperatures“, we are using it
over 20 years and magnets and people are still alive and happy.
Cheers,
Jan
> From: main_at_ammrl.groups.io <main_at_ammrl.groups.io> On Behalf Of Ulrich
> Sent: Thursday, August 29, 2024 3:20 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Tubing for N2 fill exhaust?
Hello!
The material can NOT cope with the temperatures - it is up to -5°C capable.
Sorry, please use an ordinary opaque latex tube in the correct size. So you
can vent of the cold nitrogen gas and liquid nitrogen.
We are using for the filling process a special hose system
Hi Beth,
we are using the reinforced PVC tube (softenned PVC both inner part as well
as the outer part) like this
https://www.gumex.cz/h/1125-hadice-na-pitnou-vodu-a-tekute-pozivatiny-01283?Filter=True
either 10/14mm diam. or 16/22 diam. (inner/outer) for different magnets.
This kind of tube works well, once upon a time it breaks, so we cut a new
piece. At a price of ~ 30 CZK/m it does not matter.
[NMRtools_LN2-Hose-1.png]
NMR LN2-HOSE <
https://nmrtools.com/product/nmr-ln2-hose/ >
nmrtools.com <
https://nmrtools.com/product/nmr-ln2-hose/ >
It is expensive - but it is it worth.
Yours sincerely,
Ulrich Haunz
Cheers,
Jan
---------------------------------------------------------
Jan Lang
Associate Professor
Charles University
Faculty of Mathematics & Physics
Department of Low Temperature Physics
V Holesovickach 747/2
CZ-180 00 Prague 8
Czech Republic
Tel.: +420 221912889
http://nmr.mff.cuni.cz
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1577): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1577__;!!PvDODwlR4mBZyAb0!WKzd7q0lCQfUg3XaXkUgNFbf8y6phy=
PMSD7VEIK7EHujWPc4z1QrCG7FNoDeAA7w1gDwFWGHr64c3AfhBq40tLClTIZr$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/108103713=
/7559972__;!!PvDODwlR4mBZyAb0!WKzd7q0lCQfUg3XaXkUgNFbf8y6phyPMSD7VEIK7EHujW=
Pc4z1QrCG7FNoDeAA7w1gDwFWGHr64c3AfhBq40tLMHy6f6$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Aug 30 2024 - 05:00:43 MST