We have spent some time getting C13 spectra on a cold probe in which both H1
and C13 are cold on an INOVA600. Knowing that probe ringing is substantial,
I have used the macro "coldprobe" to increase rof2 to 350us. The macro did
not seem to set up any processing procedures to deal with the data, however.
The phase seemed all over the place. I have previously done the same on a
DirectDrive system and there was no problem. I have posted this question to
Varian, but I have not gotten an answer. Does anyone know how to process
this kind of data?

 