Re: [AMMRL] inexpensive electronic flow sensors

From: Matthias Brandl via groups.io <matthias.brandl=tum.de_at_groups.io>
Date: Tue, 28 Apr 2026 16:45:40 +0200

Dear Sebastian,

Our NMR facility took delivery of three new magnets last year, and they
came with flow meters for both nitrogen and helium.

They are Vögtlin red-y compact units
(https://www.voegtlin.com/massedurchflussmesser-und-massedurchflussregler/red-y-compact-series/),
the model number of the He meter is GCM-A9EA-BN00.

They come with a factory calibration certificate and are apparently
specifically calibrated for helium. I have attached a picture of the
nameplate. Hope this helps.

Best,
Matthias



On 22/04/2026 17:45, Sebastian Kemper via groups.io wrote:
> Hello everyone,
>
> I have another question about helium recovery systems. Given the current
> situation, I would like to install more sensors in my recovery system to
> better monitor helium losses. I know many of you use rotameters to
> measure the helium flow of individual spectrometers. I would like to do
> this electronically to better calculate average values ​​and have
> everything displayed at a glance on my dashboard. I've already tried
> measuring the flow rate with a small gas meter (Honeywell BK-G1.6) with
> an electronic counter (Encoder AE5), which worked to some extent.
> However, I measured that the readings were lower than the actual flow
> rate (the flow rates of the spectrometers were just below the meter's
> specifications). Does anyone have experience with other inexpensive
> electronic sensors that measure flow rates in the range of <200 l/min?
> My internet (&AI) searches turned up results such as the SMC PFMV505-1
> and the Honeywell AWM3300.
>
> Best regards,
>
> Sebastian
>
>

--
Dr. rer. nat. Matthias Brandl (he/him or they/them)

Technical University of Munich
TUM School of Natural Sciences - BNMRZ
Chair of Biomolecular NMR Spectroscopy (Prof. Sattler)

Lichtenbergstr. 4 (postal)
Ernst-Otto-Fischer-Str. 2 (physical)
85748 Garching b. Muenchen

Phone +49 89 289-52617

matthias.brandl_at_tum.de
https://www.sattlerlab.de/
https://www.bnmrz.org/



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3418): https://ammrl.groups.io/g/main/message/3418
Mute This Topic: https://groups.io/mt/118972237/7559972
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://ammrl.groups.io/g/main/leave/12177410/7559972/1125599165/xyzzy [walt_at_hawaii.edu]
-=-=-=-=-=-=-=-=-=-=-=-




1000005891.jpg
(image/jpeg attachment: 1000005891.jpg)

Received on Tue Apr 28 2026 - 08:01:35 MST

This archive was generated by hypermail 2.4.0 : Thu Apr 30 2026 - 18:38:19 MST