Re: [AMMRL] ACB error during 19F acquisition on 400MHz prodigy probe

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Thu, 23 Apr 2026 18:08:54 +0000 (UTC)

Hi,

An ACB error is frequently a communication error and not probe related.
If you're running an older version of topspin, it might be worth updating
to the latest version. If you have a NEO console you can also try updating
the router firmware if it has a Cisco router.  It could also be an actual
ACB error--if you have an older spectrometer and you're getting this error
consistently it might be worth swapping the ACB.  If you can provide
more details about your instrument and when you're seeing the error we can
likely help more effectively.

Cheers,Mike

On Thursday, April 23, 2026 at 08:11:29 AM MDT, pothula purushotham reddy via groups.io wrote:


Hi All,

I observed ACB error during 19F acquisition, recently we installed prodigy
probe on 400MHz spectrometer.

Best regards.
Purushotham Reddy



-=-=-=-=-=-=-=-=-=-=-=-