Hi Paul,
We used to have a Fourier 300 on 3.2pl7 with a Bruker-issued HP Z420 on
Windows 7. Despite it not being officially supported, we successfully did
an in-place upgrade to Windows 10 without apparent issues. Obviously, that
doesn't guarantee it's also compatible with Windows 11, but at least there's
a reasonable chance it might.
Thanks
Geoff
________________________________
> From: main_at_ammrl.groups.io on behalf of Ralifo, Paul S. via groups.io
> Sent: 17 April 2026 18:26
> To: main_at_ammrl.groups.io
> Subject: [External] Re: [AMMRL] AVII Topspin 3.2 upgrade Windows 11
Hi Sarah
We have topspin 3.2 on windows 7 for our teaching NMR, Fourier 300 and like
you, we are being forced to upgrade. Could you possibly share with me your
computer configuration? For example your computer make and model, processor
type, RAM etc?
Thank you so much in advance.
Paul
--
Paul Ralifo Ph.D.
Department of Chemistry
Boston University
590 Commonwealth Ave, Rm B86
Boston, MA 02215
Zoom: https://bostonu.zoom.us/j/4620980251
Office: 617.353.4818
Google Voice (text preferred): 617-701-6254
ralifo_at_bu.edu
Hi all,
I am reading through old threads and trying to see if I can figure out what's
going on with our update to Windows 11 and Topspin 3.2 for an AVII system (2006).
We were previously running ok on Windows 7 with Topspin 3.1 but are being forced
to upgrade the hardware/OS. (My pleas of "don't mess with things that are working"
went unheeded.) The old computer did not have RS232 connection and it seems like
we were able to achieve everything through ethernet for 15 years, so I haven't
pursued RS232 to USB adapter. I haven't ventured down the VM route yet although
that solution seems to work for many.
I have installed the OpenText NT toolkit and Topspin 3.2. I have flexlm running.
I have restored the previous spectrometer configuration with nmr_save/ nmr_restore.
I am able to ping 149.236.99.89 without issue.
cf connects to the DRU and then hangs/errors at iiconf: connect to CCU.
"connection to spect (aqport0) failed unable to connect to remote hose
(Unknown error) ('startd' demon active on CCU ?)"
Any guidance would be appreciated. (Help me Sameer, you're my only hope, etc) :)
This is my oldest console but I have four more to go after this, unless I walk
into the sea first.
Cheers,
Sarah
Iowa State
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3395): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3395__;!!PvDODwlR4mBZyAb0!XxmiRrXFhlOMkMUAG2NWIMoqkAaAvs=
YTGyG_gdIGoTEsM3eBIZklyVS136mQFA11Av8WJr3w70-FN2yJUwus1dGUHo_T$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118692233=
/7559972__;!!PvDODwlR4mBZyAb0!XxmiRrXFhlOMkMUAG2NWIMoqkAaAvsYTGyG_gdIGoTEsM=
3eBIZklyVS136mQFA11Av8WJr3w70-FN2yJUwus1Qz0R3Oy$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Sat Apr 18 2026 - 07:59:01 MST