Re: [AMMRL] Prodigy Probe/Air Pressure

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Fri, 17 Apr 2026 15:49:52 +0000 (UTC)

Hi Simon,

Assuming the flow is consistent (and a prodigy doesn't really need that much
flow unless it's switching valves) you should probably be okay. If the pressure
drops low enough you will get error messages but I believe those points are
well below 78 psi. The regulator in the platform is supposed to be set around
82 psi so 78 really isn't that low.  You should have plenty of wiggle
room if it's just going down to 78.

Cheers,Mike

On Friday, April 17, 2026 at 09:15:51 AM MDT, Sham, Shing via groups.io wrote:

Hi,I would like to get some insights from you folks.We use a compressor air +
Bruker N2 separator for our Prodigy probe. The manual lists the pressure
requirement around ~ 100 PSI. The output of our air compressor recently
fluctuated between 100 to 78 PSI. I know it is a large fluctuation.
Has anyone tried to cool down the probe and let it run under this kind of
condition.

Thanks in advance.

Simon


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3391): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3391__;!!PvDODwlR4mBZyAb0!TEbsi2hjwWxDLOuVAIs0saYy4-IvRp=
DlwNT_S717akIQqwf68rnqA7OBC3oo_-seh205rkhM9Af0zB04X__VpgRIzNo4$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118877086=
/7559972__;!!PvDODwlR4mBZyAb0!TEbsi2hjwWxDLOuVAIs0saYy4-IvRpDlwNT_S717akIQq=
wf68rnqA7OBC3oo_-seh205rkhM9Af0zB04X__Vpi9vvoRi$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Apr 17 2026 - 09:30:59 MST

This archive was generated by hypermail 2.4.0 : Thu Apr 30 2026 - 18:38:19 MST