Re: [AMMRL] unable to run TopSpin 4.4.0 from nmr account after power outage

From: DrSpin via groups.io <rajanp=uw.edu_at_groups.io>
Date: Wed, 16 Apr 2025 09:53:38 -0700

In good old days, the 'spect' and 'ASP_ST2' definitions used to be
mandatory in /etc/hosts. When the 'eth0' and 'eth1' mappings were strictly
followed to satisfy 'flexlm', (eth0 facing the world and eth1 within the
spectrometer network), when the hostname was sometimes tied to 127.0.0.1
(linux level quirk), it used to cause some issues. Nowadays the name
resolution and IPv4 mapping has become more complex.

  In my NEO system, I do not see any such definition. This is most
probably due to two facts. One is that the IPv4 addresses of the
spectrometer network are now controlled by DHCP from the NEO console rather
than the PC box. Second, for NEO systems, we no longer have the
'diskless' download of the OS image onto the rather barebone computer (ccu,
IPSO a bit better) in the console via NFS. They have (thankfully)
rethought the whole paradigm and there is a beefed up Linux system with, I
think, 1 TB disk in the console. The bootstrap program of that system
doesn't rely on our Linux/Windows host anymore to come up and poll all the
console electronics, etc.

 If you notice, the /opt/topspin/conf/instr has 'remote_spect' rather than
'spect' for NEO onwards. In my console, the magic happens in a single
board 'SCU', as far as I could see. The data dump is in the reverse
direction i.e. the console pushes data onto the computer rather than the
controlling PC pulling it onto local disk (via the wr #0 type pulse program
statements). Of course, if it's data 'pull' or 'push' is a bit academic.
But, this standalone system that is in NEO console allows us to take some
nifty steps such as restarting the console electronics 'remotely' from
within Topspin (gone are the days of pushing the Red button on CCU or IPSO
i.e. you have to physically get to the lab on a Sunday afternoon, when
something is hanging).

I also will note that, when we upgrade the version of TS, the 'reboot' and
upload of the necessary server components (onto the console electronics)
now happens during the installation itself. Cool upgrade !

Rajan

On Tue, Apr 15, 2025 at 7:24 PM Michael Groves via groups.io wrote:

> You could also try deleting the C:\Users\<username>\.topspin-<pcname>\
> folder. I believe this is another recommended fix when you get these
> 127.0.0.1 error messages. This does assume you're running Windows--I'm not
> quite sure where that folder lives on a Linux system.
>
> Cheers,
> Mike
>
> On Tuesday, April 15, 2025 at 05:20:35 PM MDT, Harwood, John S via groups.io wrote:
>
>
> Tom,
>
> We had this issue crop up on one of our NEO500s running TopSpin 4.3. One
> user out of many couldn't run TopSpin with the same symptoms you describe.
> I (apparently) fixed it by deleting the user's curdir entry. If you try
> this make sure first that there are no TopSpin processes running before you
> delete the directory.
>
> Sincerely,
>
> JSH
>
> John S. Harwood, D.Sc., Director
> The Robert Santini NMR Facility at Purdue University
> http://www.pinmrf.purdue.edu
> jharwood_at_purdue.edu
> 1-765-494-5287
> ------------------------------
> *From:* main_at_ammrl.groups.io on behalf of Tom Pratum via groups.io
> *Sent:* Tuesday, April 15, 2025 4:28 PM
> *To:* main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> *Subject:* [AMMRL] unable to run TopSpin 4.4.0 from nmr account after power outage
>
>
>
> We had a power outage over the weekend and after that we have been only
> able to run TopSpin from numrsu. Even if both nmr and nmrsu are logged in,
> TopSpin cannot be run from the nmr account and gives a long list of Java
> related errors (which do not appear from nmrsu), one of which is a
> connection related exception to 127.0.0.1:8080. Has anyone seen this
> previously? Any ideas as to what needs to be ‘reset’ in
> order to get past
> it?
>
> Thanks in advance for any help.
>
>
>
> Tom Pratum, Southern Oregon University
>

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2239): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2239__;!!PvDODwlR4mBZyAb0!ShVLZTCclB3mdlqxBSTqjUiflC41CO=
-QOT5zABq7qwqcZChzNO5qHpFiFbWNcXZFHamCwoiopZLcSnaB4Yb3tPX5$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/112284157=
/7559972__;!!PvDODwlR4mBZyAb0!ShVLZTCclB3mdlqxBSTqjUiflC41CO-QOT5zABq7qwqcZ=
ChzNO5qHpFiFbWNcXZFHamCwoiopZLcSnaB4SBUUl9Q$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Apr 16 2025 - 10:02:35 MST

This archive was generated by hypermail 2.4.0 : Thu May 01 2025 - 13:19:57 MST