Re: [AMMRL] unable to run TopSpin 4.4.0 from nmr account after power outage

From: Tom Pratum via groups.io <tkp9551=gmail.com_at_groups.io>
Date: Thu, 24 Apr 2025 10:35:32 -0700

I wanted to post that removing the <TSHOME>\prog\curdir\<user>\nmrdata-search
folder for the user involved solved this issue. Some form of this solution
was suggested by a couple of folks who responded here, and also eventually
by Bruker.

Thanks for all of your help!

Tom Pratum


> On Apr 15, 2025, at 1:28 PM, Tom Pratum via groups.io wrote:
>
> We had a power outage over the weekend and after that we have been only able
> to run TopSpin from numrsu. Even if both nmr and nmrsu are logged in, TopSpin
> cannot be run from the nmr account and gives a long list of Java related
> errors (which do not appear from nmrsu), one of which is a connection
> related exception to 127.0.0.1:8080. Has anyone seen this previously? Any
> ideas as to what needs to be ‘reset’ in order to get past it?
>
> Thanks in advance for any help.
>
>
>
> Tom Pratum, Southern Oregon University
>



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2272): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2272__;!!PvDODwlR4mBZyAb0!Q9x0x-e-K31Et20MghM1evYrXfabvs=
vv4_gG_4arFudR5XCy1_epVf8SCY9x9a7XvOS5nmBW1N-ixpgyB1sK6pHgNm09$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/112284157=
/7559972__;!!PvDODwlR4mBZyAb0!Q9x0x-e-K31Et20MghM1evYrXfabvsvv4_gG_4arFudR5=
XCy1_epVf8SCY9x9a7XvOS5nmBW1N-ixpgyB1sK6gBgvAxd$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-

Received on Thu Apr 24 2025 - 11:06:43 MST

This archive was generated by hypermail 2.4.0 : Thu May 01 2025 - 13:19:57 MST