Hi,
recently a friend had a nice challenge.
His equipment
- Avance III console
- BSMS/2
- B-VT 3200 as part of the BSMS/2
Since a few days he get's an overheating message. Switching cable and heater
doesn't help, neither switching the whole B-VT3200 rack.
Maybe the extra 48V power supply next to the B-VT 3200 at the rear side of the BSMS/2
might be the reason. An indication is an additional - non disturbing - error message about
a failing nitrogen evaporator.
Of course that friend could change the power supply on Monday. Nevertheless I am very curious.
Did you ever hear about the combination
failing 48V power supply - "overheating" message in the past?
Greetings
Rainer
-=-=-=-=-=-=-=-=-=-=-=-