Re: [AMMRL] Varian VT Connector Fix

From: Indrek Reile <indrek.reile_at_kbfi.ee>
Date: Mon, 7 Apr 2025 17:50:59 +0300

Hey,

That's a very familiar problem. At some point half of our guys in the
lab developed the skill to resolder a broken wire to the connector in a
couple of minutes.

Finally someone here had the idea of suspending the VT cable with a rope
that is hanging from the magnet rim. That way the connector does not
carry the whole weight of the cable, most of the strain is held by the
suspending rope. We have not had a single connector breakdown since.

Regards,

Indrek

---
Indrek Reile, Ph.D.

Senior Research Fellow / Solution NMR
Member of Science Council
National Institute of Chemical Physics & Biophysics
Akadeemia tee 23, 12618 Tallinn, Estonia
email:indrek.reile_at_kbfi.ee
skype: indrek.r
mobile: +372 523 6054

03.04.2025 1:10 Casey Simons via groups.io kirjutas:

> Hey all! I wanted to check with the spinlanders before trying to
> re-invent the wheel.  Has anyone developed a replacement connect for
> the Varian VT serial connector?  I'm running both a VNMRS and a INOVA
> consoles and looking for a more reliable connection for the VT sensor.
>
> C
>
> Casey Simons
> NMR Director
> CAMCOR Nanofabrication and Imaging Facility
> University of Oregon


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2215): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2215__;!!PvDODwlR4mBZyAb0!Qv1OFXbNvFVIQe4I161xK0JGC2lcLk=
jby3mLS0vx_fkUTjBafyJbsl5qfu-6bjRPm3J-dL1bv2CDZ3GGjmjegQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/112057683=
/7559972__;!!PvDODwlR4mBZyAb0!Qv1OFXbNvFVIQe4I161xK0JGC2lcLkjby3mLS0vx_fkUT=
jBafyJbsl5qfu-6bjRPm3J-dL1bv2CDZ3H0FqBmLw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Apr 07 2025 - 16:57:44 MST

This archive was generated by hypermail 2.4.0 : Thu May 01 2025 - 13:19:57 MST