Re: [AMMRL] Varian VT Connector Fix

From: Olson, Dean via groups.io <dolson=illinois.edu_at_groups.io>
Date: Thu, 3 Apr 2025 14:59:35 +0000

Casey and Colleagues,

We took apart all those connectors and made sure all the wires were soldered
to the pins. That solved most problems most of the time.

So, we did not replace the connectors, we improved the connectors. We have
not had the +/_999.9 error (a communication error) in a very long time.

Dean

***********************************
Dean L. Olson, Ph.D.
Director, NMR/EPR Lab
146 Roger Adams Lab
School of Chemical Sciences
U. of Illinois at Urbana-Champaign
600 S. Mathews, Box 31-1, MC-712
Urbana, IL 61801 USA
Lab: 217-244-0564; Cell: 217-722-9432
dolson_at_illinois.edu
http://scs.illinois.edu/nmr
http://scs.illinois.edu/epr/
***********************************

> From: main_at_ammrl.groups.io On Behalf Of Casey Simons via groups.io
> Sent: Wednesday, April 2, 2025 5:11 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Varian VT Connector Fix

Hey all! I wanted to check with the spinlanders before trying to re-invent
the wheel. Has anyone developed a replacement connect for the Varian VT serial
connector? I'm running both a VNMRS and a INOVA consoles and looking
for a more reliable connection for the VT sensor.

C

Casey Simons
NMR Director
CAMCOR Nanofabrication and Imaging Facility
University of Oregon



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2214): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2214__;!!PvDODwlR4mBZyAb0!XYjAgfOSlCA4wwmUCIRJww04wMhKzb=
qPlk3y6t0gSviNeCbnUQs5WhruZQegBJjCMbl8q8NLyN1aU46CPaIkctzKo1Uv$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/112057683=
/7559972__;!!PvDODwlR4mBZyAb0!XYjAgfOSlCA4wwmUCIRJww04wMhKzbqPlk3y6t0gSviNe=
CbnUQs5WhruZQegBJjCMbl8q8NLyN1aU46CPaIkcivcTwu0$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-



Received on Mon Apr 07 2025 - 07:19:39 MST

This archive was generated by hypermail 2.4.0 : Thu May 01 2025 - 13:19:57 MST