 Hi Alex,
Hopefully smarter people than me will chime in and have a good answer for this. 
In the meantime:

What OS is this on?

Does this happen for all users?  It's happened that somebody made a username
with a space in it and that caused java some headaches.

Has your network configuration recently changed?  Apparently if your hosts file
(C:\Windows\System32\drivers\etc\hosts, or /etc/hosts) doesn't match up with the
current network or is missing something then it can cause problems.  I believe
the issue that came up was because the local hostname wasn't in the hosts file
The fix was to add a line to the hosts file with the syntax "127.0.0.1 <hostname>"

One other person ran into this and had to change an environment variable--they
said the following:Premature termination of java virtual machineProgram is exiting
...Hit ENTER to continue ...- Reason: no idea.- Workaround:- Solution:Just add a
"system variable" in the control panel, system properties.The variable I added was
"_JAVA_OPTIONS", with the value "-Xmx512m".See
http://help.dugeo.com/m/faq/l/98932-how-do-i-edit-the-java-s-overriding-environment-variable*22for
a fuller explanation.Reinstalling TopSpin does not help.

Anyway, that's what I've got.  Good luck.

Cheers,Mike

    On Thursday, April 25, 2024 at 11:19:20 AM MDT, Blumenfeld, Alexander wrote:

Dear all,
This is from Alex Blumenfeld, University of Idaho, with Bruker Avance III
500 MHz, TopSpin 3.2. We lost access to the TopSpin environment. Each
time we click on theTopSpin icon, the following error message shows up:

Premature Java virtual machine termination. The program is exiting...Hit Enter to continue

Obviously, hitting Enter just closes this window.

Our IT specialist checked all memory allocations and did not find any overfills.

Any advise will be appreciated!

Thank you,

Alex Blumenfeld
NMR Lab Manager
University of Idaho
Moscow ID,208-892-2437



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1191): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1191__;!!PvDODwlR4mBZyAb0!U_aCc900ixmzg9xfkE5Mhr_qkT8Ssf=
NvDbnFByClzW9bMRQPqkZbbTWfphpbwQH7AgPjBaDjJIomiFn56TOzXoIn3-DL$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105734864=
/7559972__;!!PvDODwlR4mBZyAb0!U_aCc900ixmzg9xfkE5Mhr_qkT8SsfNvDbnFByClzW9bM=
RQPqkZbbTWfphpbwQH7AgPjBaDjJIomiFn56TOzXmbpD4db$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Apr 25 2024 - 12:02:38 MST

This archive was generated by hypermail 2.4.0 : Wed May 01 2024 - 15:08:18 MST