Hello,

that's no direct answer concerning your question.

If you have really sensitive samples, the best way is to add a tiny amount
silver oxide directly into the NMR tube. You might be interested to check out the effect.

Take CDCl3 as a solvent and add a drop of methanol. You should see two singlets, one for
the methyl, one for The OH group. Now add a few micrograms of silver oxide. You have to
shake it a rather long time. Long time means At least one hour. An ultrasonic bath should do the job.
Insert the sample immediately into the spectrometer to keep light away and wait a few minutes to allow the silver
Oxide to move to the bottom of the tube.
And then measure again. You should see the expected doublet for the Methyl and a quartet for the OH group.

Greetings

Rainer



Von: main_at_ammrl.groups.io <main_at_ammrl.groups.io> Im Auftrag von Vernon M Russell
Gesendet: Dienstag, 23. April 2024 20:53
An: main_at_ammrl.groups.io
Betreff: [AMMRL] Silver Strips in CDCl3?

Has anyone tried adding a 'silver strip' to a bottle of CDCl3 for stabilization?
Does anyone know where to buy silver strips?
Thanks,
Vernon

Vernon Russell
Analytical Team / Gore Incubator
vrussell_at_wlgore.com<mailto:vrussell_at_wlgore.com>
T +1-410-506-3852

W. L. Gore & Associates, Inc.
Together, improving life
501 Vieve's Way, P.O. Box 1320
Elkton, MD 21922-1320
USA


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1188): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1188__;!!PvDODwlR4mBZyAb0!UZjLZXpz1hVHjNOZthRpLthAjyjwhR=
8IcJUYEYO5ouHmcW_sHTOfYc4S6gQ3CGnkVge4bHw-_w1f-g$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105695816=
/7559972__;!!PvDODwlR4mBZyAb0!UZjLZXpz1hVHjNOZthRpLthAjyjwhR8IcJUYEYO5ouHmc=
W_sHTOfYc4S6gQ3CGnkVge4bHxNs9b5zw$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Apr 24 2024 - 00:44:34 MST

This archive was generated by hypermail 2.4.0 : Wed May 01 2024 - 15:08:18 MST