Hi John,
The au programs 'savepars' and 'getpars' can be used to do this. I've done
this for some of mine by executing these au programs from within the
program in my AUNM field with the lines* XCMD("savepars")* and
*XCMD("getpars")*. There may be other and/or better ways to do this, but
this is the solution we've been using.
Best,
Peter
On Tue, Apr 9, 2024 at 12:35 PM John H Grimes Jr. via groups.io wrote:
> When acquiring data in IconNMR we often call pulsecal in the AUNM field.
> Rather than having to call it for each experiment of a given sample, we
> would like to call it for the first experiment (say a proton), and then
> have that result copied to the other experiments that are acquired from
> that sample. I have browsed my manuals but not come across this function.
> Does this capability exist and I am just missing something?
>
> Thanks,
>
> John
>
> John Grimes | NMR Facility Manager
> Complex Carbohydrate Research Center, University of Georgia
> 315 Riverbend Road
> Athens, GA 30602-4712
> (706) 542-4483
> jgrimesjr_at_uga.edu
>
>
>
>
--
*Peter Thompson, Ph.D.*
Lead Research Scientist, Molecular Education, Technology and Research
Innovation Center (METRIC)
Research Assistant Professor, Department of Molecular and Structural
Biochemistry
North Carolina State University
51 Polk Hall
Raleigh, NC 27695
Phone: 919.515.0836
Email: pmthomps_at_ncsu.edu
METRIC Website: https://research.ncsu.edu/metric/home
Consultation and Sample Submission: https://labs.ncsu.edu
Biochemistry Website:
https://cals.ncsu.edu/molecular-and-structural-biochemistry/
“All electronic mail messages in connection with State business which are
sent to or received by this account are subject to the NC Public Records
Law and may be disclosed to third parties.”
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1117): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1117__;!!PvDODwlR4mBZyAb0!XOxBhAASTEZNz3tGs5RB6FRniDHVjv=
VcucN4FmF4nyEX-I8NQxRyonS-n5aGWA-Md6j5h8MReoIpz7fE1TuiIESVx_RJ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/105425381=
/7559972__;!!PvDODwlR4mBZyAb0!XOxBhAASTEZNz3tGs5RB6FRniDHVjvVcucN4FmF4nyEX-=
I8NQxRyonS-n5aGWA-Md6j5h8MReoIpz7fE1TuiINEouTut$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Apr 09 2024 - 09:45:02 MST