Re: [AMMRL] trouble in helium recovery land...

From: Michael Groves via groups.io <mgrovesnmr=yahoo.com_at_groups.io>
Date: Thu, 2 Apr 2026 20:06:15 +0000 (UTC)

Sounds like you've done a pretty thorough job.  Could there be a valve
on the EPR or MCD that's leaking while that equipment isn't connected? 
Seems like if it was left open all the way you'd get a bunch of air in the
system but maybe it's not sealing properly when closed?  Have you opened
up the compressor?  There are probably a couple of connections in there
as well--you might see if the compressor is leaking internally.  

Cheers,Mike

-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3341): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3341__;!!PvDODwlR4mBZyAb0!Xlzq9fYhs_dWL0DNYDS3QAaAiG1CLL=
UlT04bAEQVgprw0PFpHxSrcwEoU3NE8sGy917bcc798w7RYNZuLi3GYVulw8XD$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118639575=
/7559972__;!!PvDODwlR4mBZyAb0!Xlzq9fYhs_dWL0DNYDS3QAaAiG1CLLUlT04bAEQVgprw0=
PFpHxSrcwEoU3NE8sGy917bcc798w7RYNZuLi3GYc1x4MNh$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Apr 03 2026 - 07:38:25 MST

This archive was generated by hypermail 2.4.0 : Sun Apr 19 2026 - 19:14:00 MST