Re: [AMMRL] Noise complaints from Quantum Technology Helium Recovery System

From: Fulton, Luke via groups.io <fko24003=uconn.edu_at_groups.io>
Date: Thu, 2 Apr 2026 18:47:46 +0000

Hi Xu,

I actually like the noise, those chirps reports on cold head operations, an d
their absence is an immediate alarm to investigate. I wouldn't recommend
eliminating the noise entirely were it possible. Our dewar has a soft wrapping
around it to dampen the volume, from touch/appearance I'd say it's all
synthetics and foam. We sourced it directly from QT as part of the original
install and would assume it's purchasable as a stand alone accessory. You
could probably assemble something custom yourself to achieve similar results.

I haven't noticed the chirping volume being any more severe than for helium
cooled cryoprobes. Users might find it annoying from personal preference,
but it doesn't rise to safety concern in my opinion. The HR3 compressor on
the other hand is much louder. Thankfully it only turns on during high boil
off scenarios which are either rare or predictable such as fills. I suggest/offer
ear protection to users working long experiments while it's running.

Kind regards,
Luke


Luke Fulton, PhD
CHEM BLDG R003
NMR Core Facility Director
Unit 3060
COR2E & Department of Chemistry
55 N Eagleville Road
University of Connecticut
Storrs, CT 06279

email: fko24003_at_uconn.edu
(alias): luke.fulton_at_uconn.edu
mobile: (603) 953-5275
Office: (860) 486-4069
________________________________
> From: main_at_ammrl.groups.io on behalf of Xu, Xu via groups.io
> Sent: Tuesday, March 31, 2026 3:23 PM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] Noise complaints from Quantum Technology Helium Recovery System

Hi All,

Anyone has complaints about the chirping noise from QT Helium Recovery System?
How do you deal with this?

Best,
-Xu



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3326): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3326__;!!PvDODwlR4mBZyAb0!TQX39mX6H5enIXb5cjW2QT0JrzSdUS=
XSVQZJBO42Gxh9wHX9gPbLCxQzxDuPumxiCSWof4yIiyPF1SM8mKndyQu0Rhxx$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/118637955=
/7559972__;!!PvDODwlR4mBZyAb0!TQX39mX6H5enIXb5cjW2QT0JrzSdUSXSVQZJBO42Gxh9w=
HX9gPbLCxQzxDuPumxiCSWof4yIiyPF1SM8mKndyVKM9AnV$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Thu Apr 02 2026 - 11:52:50 MST

This archive was generated by hypermail 2.4.0 : Sun Apr 19 2026 - 19:14:00 MST