Hi Sam,
Your SampleXpress should've originally come with a blue case that comes with
all sorts of spares, including different lengths of all of those parts.
The threaded "struts" can be screwed in or out at feet at the bottom and in
to the body at the top to adjust the overall length. The top part of the
SampleXpress has lots of space inside so maybe it just got screwed all the
way in and never restored back to rights. The feet are separate parts from
the threaded studs so you could replace them with literally anything provided
they aren't magnetic.
Similarly, the lower parts of the "cage" are swappable to different lengths,
and also height adjustable using the clamping screws on each one. They are
a bit more complicated, so if you don't have the spares box maybe the simplest
answer is to get your mechanical workshop to fashion some new ones. The bolted
collars come off and they can be removed from the bottom ring (see the wrench
flats), so they are mechanically quite straightforward to copy. On our US400+SX
we are using the longest versions of these.
Or ask Bruker UK nicely to see if they have some spares lying around.
If none of that makes sense email me directly and I'll send you pictures to
clarify.
Thanks
Geoff
________________________________
> From: main_at_ammrl.groups.io on behalf of Samuel Digby via groups.io
> Sent: 16 September 2026 19:32
> To: main_at_ammrl.groups.io
> Subject: [External] [AMMRL] SampleXpress mounting rings
Hi all,
I don't suppose anyone would have any SampleXpress mounting rings they have
no intention of ever using again and would be willing to part with for a
suitable price? And possibly any of the longer struts that support the front
of the unit against the magnet (for some reason this thing has one long and
one short)?
Trying to fit this to an ultrashield 400 magnet and the stack is a little bit
too long, so I need the carousel to sit higher than it does using the one I
have (originally fit to an ascend 400 where the stack pokes out a bit less).
For avoidance of doubt I've attached an image.
Thanks,
Sam
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3790): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3790__;!!PvDODwlR4mBZyAb0!Xx050OkCDCRpzRhExLUWVA7JFxWPui=
_6gZhhfn96gxvrnSa63Se4x2cRtwbdqDWJAo-zfFNSSnSM0KDgDmPImhEM37OR$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/121283816=
/7559972__;!!PvDODwlR4mBZyAb0!Xx050OkCDCRpzRhExLUWVA7JFxWPui_6gZhhfn96gxvrn=
Sa63Se4x2cRtwbdqDWJAo-zfFNSSnSM0KDgDmPImrsi8yBR$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Thu Sep 17 2026 - 04:56:22 MST