Re: [AMMRL] Impact testing for nearby construction?

From: Laurie Galya - EXT via groups.io <lgalya=preludetx.com_at_groups.io>
Date: Mon, 15 Sep 2025 19:42:30 +0000

Hi Beth,

When I was at Incyte, they built a new building relatively close to the research
building. The new building required blasting. The engineers installed an
instrument like a seismometer in the NMR lab. They did a couple of test blasts
to get their parameters set before the actual blast and we could see what kind
of problems we were going to have.

We had anti-vibration legs and really had no problems at all. The plane of
some of the blasts were a bit above the plane of the rock that the research
building was siting on. See if you can find out if they are going to do
any kinds of tests to give you an idea of what may happen to your systems.

Good luck!
Laurie

> From: main_at_ammrl.groups.io On Behalf Of Beth Moscatovia groups.io
> Sent: Monday, September 15, 2025 3:08 PM
> To: main <main_at_ammrl.groups.io>
> Subject: [EXTERNAL]:[AMMRL] Impact testing for nearby construction?


Hi all

We recently learned that a wing of our building will be demolished next summer.
Several people in facilities (none of them in chemistry) have told the department
chair that the demolition will not impact the research wing in any way. It’s
unclear how they have decided this, especially since – as far as I can tell –
no experimental tests have been done to see whether any demolition work will carry
through the slab to nearby parts of the building.

Our magnets do not have anti-vibration legs, so we are particularly vulnerable to
any nearby work. I know that the laser labs are also concerned.

I’ve searched through AMMRL and found several excellent writeups on
things to consider when buildings may be under construction, but I haven’t
seen any discussion about what kinds of tests we could do to see how sensitive the
magnets will be to any sort of construction work. I know that 2D experiments are
the most susceptible to disruption, but are there any recommended protocols?
Similarly, what sorts of construction work (jack hammering, etc.) should I be most
concerned about? Are there any experimental tests I can request on the facilities
side to predict in advance what sorts of problems we might have?

Does anyone have any pointers about specific data that I should expect facilities
to be able to provide that would show that they have done the work to determine
what potential vibrational impacts construction would cause?

It does not seem like we will be able to prevent the construction from occurring but
even knowing in advance what parts of the demolition will create the most problems
so that students can schedule experiments around those times will be valuable.

One writeup suggested paying for an external consultant to scout out the labs and
identify possible issues. Assuming facilities has not done due diligence, we can
probably do that, but I would like to know what concerns I should raise in order
to justify doing so.

Thank you!
Beth Moscato
Temple University



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2809): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2809__;!!PvDODwlR4mBZyAb0!ST0hilhpLcqJMkyG4zkgBUaFMiLe83=
YnJcrT2GhTuB0H5Oqk7lC4EOSIFy1RwMiLyhNdZ_65eBqWPbP5uZP1LZFUohLy$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/115260442=
/7559972__;!!PvDODwlR4mBZyAb0!ST0hilhpLcqJMkyG4zkgBUaFMiLe83YnJcrT2GhTuB0H5=
Oqk7lC4EOSIFy1RwMiLyhNdZ_65eBqWPbP5uZP1LTpOv4wL$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Sep 15 2025 - 12:51:56 MST

This archive was generated by hypermail 2.4.0 : Wed Oct 01 2025 - 14:23:05 MST