some hint for the next refill: touch the helium transfer line along the length
line during and after helium filling:
- if it feels well below ambient temperature during refill, the vacuum inside is
really bad, needs pumping (it's normal to be cold around the two bends though,
even a small frozen spot may happen)
- if it gets quite cold after refill (in bad cases you may see some rings forming
where condensation is less, every 15-20 cm or so - these are the internal spacers),
then it's not as bad but re-pumping would be recommended
Other than that I concur with Paulo - doing the transfer slowly can easily
take much longer, plus suffering from increased losses (longer contact time
of the liquid helium in the "warm" transfer line)…
Cheers
Bert
> Am 09.09.2025 um 21:26 schrieb Thomas Eubanks via groups.io:
>
> You have to push the helium fast enough to stay liquid into the magnet.
> Depending on the transfer lines you might have to go up to 3-4 lbs psi. New
> lines with strong vacuum can transfer much lower.
> From: main_at_ammrl.groups.io behalf of falcocobra via groups.io
> Sent: Tuesday, September 9, 2025 1:09 PM
> Subject: Re: [AMMRL] How to diagnose helium icing
>
>
>
> At what pressure is the fill being done? 60-80L taking more than an hour
> does sound a little slow, but if he is doing the fill at 1 psi, that doesn't
> sound out of the ordinary.
> --
> =
> Paulo Falco Cobra | Scientist II - Facility Manager
> NMRFAM <https://nmrfam.wisc.edu/> - University of Wisconsin Madison
>
> E paulo_at_nmrfam.wisc.edu <mailto:paulo_at_nmrfam.wisc.edu>
> P (608) 265-3303
> A 433 Babcock Drive, Madison, WI. 53706
>
Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist
Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net
Representing:
Stelar Fast Field Cycling NMR Relaxometers (
http://www.stelar.it
MestreLab NMR Software (
http://www.mestrelab.com
Barthel HF Amplifiers (
http://www.barthel-hf.de
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2797): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2797__;!!PvDODwlR4mBZyAb0!Vcnc8cgpr7SIGMLSbtvGRQDmBo4WBu=
ao45vka3SNW3VByu_TwCPwCaoC5pafuhEovvWGEmJQCgOOqEn5XkkIcA$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/115148493=
/7559972__;!!PvDODwlR4mBZyAb0!Vcnc8cgpr7SIGMLSbtvGRQDmBo4WBuao45vka3SNW3VBy=
u_TwCPwCaoC5pafuhEovvWGEmJQCgOOqEmUh0Kq7A$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Sep 09 2025 - 14:21:47 MST