Re: [AMMRL] AMMRL: LHe leaks - Oxford Magnet #HeliumRecovery

From: Sham, Shing <ssham_at_luc.edu>
Date: Mon, 30 Sep 2024 16:04:20 +0000

Hi Hla,

Good morning.

When was the last time you had changed those O-rings, given the age of your
magnet?

If you feel comfortable of doing that, you could change them to see whether
it wiould fix the leaks.

Hope this helps.

Simon


> From: main_at_ammrl.groups.io On Behalf Of Hla Win via groups.io
> Sent: Monday, September 30, 2024 10:42 AM
> To: main_at_ammrl.groups.io
> Subject: [AMMRL] AMMRL: LHe leaks - Oxford Magnet #HeliumRecovery

Hello Everyone,

I hope all are well on your side. We have a 500MHz Oxford 22-year-old magnet
that is leaking LHe. Picture attach. The normal LHe boil-off is 5 cc/h, and
the average loss is 0.43%, about 23 mL/h. LHe fill is every 90-100 days. It
uses about 45-50L the most. We've lost LHe this week 4x higher than normal.
You will see bubbles around the base, by the transfer port, by the little "black"
seal. If you know of anyone who repair magnet, please let me know. In the
meantime, I would appreciate any recommendations on what I should do to minimize
the loss.

Thank you.

Hla Win
Organic & NMR Lab Coordinator
Department of Chemistry https://chemistry.wwu.edu/
Western Washington University
516 High Street, Bellingham WA 98225
winpiah_at_wwu.edu<mailto:winpiah_at_wwu.edu> | 360-650-2261<tel:360-650-2261>



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1665): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1665__;!!PvDODwlR4mBZyAb0!TjnUGdQ_JtrjqX3yS8y_RlwWX8M79N=
sWCkipfFrywBxA4l_NZ8iHfArTpYHp0Fx7gddaBw8T2xNh$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/108738789=
/7559972__;!!PvDODwlR4mBZyAb0!TjnUGdQ_JtrjqX3yS8y_RlwWX8M79NsWCkipfFrywBxA4=
l_NZ8iHfArTpYHp0Fx7gddaB3nG-QFe$
Mute #heliumrecovery:https://urldefense.com/v3/__https://ammrl.groups.io/g/=
main/mutehashtag/heliumrecovery__;!!PvDODwlR4mBZyAb0!TjnUGdQ_JtrjqX3yS8y_Rl=
wWX8M79NsWCkipfFrywBxA4l_NZ8iHfArTpYHp0Fx7gddaBxFofHcX$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Sep 30 2024 - 09:06:16 MST

This archive was generated by hypermail 2.4.0 : Tue Oct 01 2024 - 14:06:56 MST