Re: [AMMRL] BBFO probe won't tune

From: Anjaneyulu Putta <anjaneyuluputta_at_boisestate.edu>
Date: Fri, 27 Sep 2024 16:22:07 -0600

Hi Beth,

Couple of years ago, we had a similar problem on tuning the X channel while
1H tuning and matching worked ok on an iprobe. So we sent this iprobe to
Bruker, there it worked well without any issues and later we found out that
a software update fixed the tuning issue. See if you have to do any
software update in the service tool section.

Best,
Anjaneyulu

On Fri, Sep 27, 2024 at 3:42 PM Beth Moscato via groups.io wrote:

> Hi –
>
>
>
> We have a 400MHz Bruker room-temp BBFO probe which has stopped tuning on
> the 1H channel. ATMA can’t find a dip, and I don’t see a change in the
> profile upon adjusting the tune manually using atmm (let alone a dip). The
> probe tunes and matches fine on the X-channel, and adjusting the match on
> the 1H channel changes the profile I’m detecting (although still, no dip).
> This is true for several samples of different solvents (CDCl3, DMSO, D2O).
> X-channel is tuned to 13C, so it’s not a problem of 19F detuning the 1H
> channel.
>
>
>
> A backup probe tunes and matches just fine, so the problem is clearly the
> 1H tuning.
>
>
>
> I have two questions:
>
>
>
> 1. Is there anything we can do that might **easily** fix this
> in-house? It seems like either the tune rod is jammed somehow and won’t
> twist (which we might be able to easily fix) or that the screws have been
> stripped (more difficult repair and probably not worth attempting).
> 2. Does anyone have third party vendors they would recommend for probe
> repairs? We are in the Philadelphia area, if that matters.
>
>
>
> Thank you guys for any advice!
>
> Beth Moscato
>
> Temple University
>
>
>

--
Anjaneyulu Putta, Ph.D.
NMR Manager
Department of Chemistry & Biochemistry
College of Arts & Sciences
Boise State University
1910 University Drive, Boise, Idaho, 83725
Phone: (208) 426-4912
http://chemistry.boisestate.edu/
https://www.boisestate.edu/chemistry/nmr-service-center/


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1659): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1659__;!!PvDODwlR4mBZyAb0!UO_iMXi2ZTWWqIuzOEhzxJ5SoZvJOA=
f7wX6wi0HGdPVXXR7oYT87T7s6gTtoIXBRU2Ok0mN49jjnFBPWpTyLToomABSVRcc$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/108695103=
/7559972__;!!PvDODwlR4mBZyAb0!UO_iMXi2ZTWWqIuzOEhzxJ5SoZvJOAf7wX6wi0HGdPVXX=
R7oYT87T7s6gTtoIXBRU2Ok0mN49jjnFBPWpTyLToomWEdgfZA$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Sep 27 2024 - 15:43:28 MST

This archive was generated by hypermail 2.4.0 : Tue Oct 01 2024 - 14:06:56 MST