Re: [AMMRL] BACs downwards motion failed

From: Gopal Subramaniam <gopal.subramaniam_at_qc.cuny.edu>
Date: Wed, 25 Sep 2024 20:50:14 +0000

We had this issue last year.

We tried several things like cable change, o-rings, cleaning, resetting boards,
etc. Replacing the arm fixed the problem and nothing else helped. It was not
intuitive to think about the arm but looks like the issue is somewhere in there.

Best

Gopal Subramaniam
Dept of Chem & Biochem
Queens College CUNY
Flushing, NY 11367

________________________________
> From: main_at_ammrl.groups.io on behalf of Justin Douglas
> Sent: Wednesday, September 25, 2024 3:25 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: [AMMRL] BACs downwards motion failed


Hello,

Our BACs autosampler (on a Bruker AV400) is producing a "DOWNWARDS MOTION FAILED"
error, which disrupts the queue in ICON (and leaves the lift air on, sometimes
all night!). The advice I was given by Bruker was to clean the arm. It is not
100% clear to me what specifically needs to be cleaned. In addition, we are
getting this error more and more frequently and I'm having to clean the arm more
and more often. It is not clear that it is really that dirty!

My method is to take special care to remove the arm, remove the bolts with an allen
wrench and wipe the piston thoroughly with chemwipes and methanol. I also use q-tips
and methanol to swab the o-rings, etc. I'm wondering if anyone has any more specific
advice on how to stop (or at least decrease the frequency of) these DOWNWARD MOTION \
FAILED errors.

Thanks

Justin


Justin Douglas
he/him/his
NMR core lab
University of Kansas
justindouglas_at_ku.edu
justindo_at_gmail.com


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1642): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1642__;!!PvDODwlR4mBZyAb0!WIvPjmBYPDTkBhsxu9tBX9Q3Ff4WJz=
EahsSgpIGmCWHy_fLe-xGowFxFl9R2DnMKSngDB5KEb-Helxu2KOu-XF6AneM0OR8$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/108655772=
/7559972__;!!PvDODwlR4mBZyAb0!WIvPjmBYPDTkBhsxu9tBX9Q3Ff4WJzEahsSgpIGmCWHy_=
fLe-xGowFxFl9R2DnMKSngDB5KEb-Helxu2KOu-XF6AQM0L3Sc$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Wed Sep 25 2024 - 13:50:20 MST

This archive was generated by hypermail 2.4.0 : Tue Oct 01 2024 - 14:06:56 MST