The tubing you most likely want is HDPE tubing. This has the lowest permeability
to moisture and CO2 and can handle pressures and temperatures that the system needs.
A problem in ordering this kind of tubing is that sellers often just say they are
selling "polyethylene tubing", leaving some doubt as to whether they are selling
the HDPE you want or (much more likely) LDPE/LLDPE that you don't. Avantor offers
HDPE in metric (catalog MFLX95501-04 ).
-=-=-=-=-=-=-=-=-=-=-=-