Re: [AMMRL] Vibration and magnet safety

From: Thomas K Weldeghiorghis <thomaskw_at_lsu.edu>
Date: Sun, 22 Sep 2024 21:38:45 +0000


Dear all,

Thank you for your responses. According to everyone's feedback, vibrations
from construction work 100 ft away won't affect NMR magnets.

Thank you,
Thomas


Thomas K. Weldeghiorghis, Ph.D.
Director, NMR Facility
Department of Chemistry
Louisiana State University
232 Choppin Hall
Baton Rouge, LA 70803
225-578-5346

________________________________
> From: main_at_ammrl.groups.io on behalf of Martha Morton via groups.io
> Sent: Friday, September 20, 2024 4:02 PM
> To: main_at_ammrl.groups.io <main_at_ammrl.groups.io>
> Subject: Re: [AMMRL] Vibration and magnet safety

Thomas,

You could check for vibrations like GE NMR engineers did; set a cup of coffee
on the floor and look for ripples. I concur with Gary, but the gross vibration
test will put your mind at ease.

Martha

On Fri, Sep 20, 2024 at 2:12 PM Thomas K Weldeghiorghis via groups.io wrote:

Dear all,

We have a water leak outside the building housing three NMR magnets, and our
university Facility Services have to demolish concrete slabs (stairs) to
fix the leak. They used an excavator breaker to break part of the concrete,
creating a vibration in the building. The construction site is about 80ft
and 100 ft from two magnets with ANTIVIBRATIONFEET TYP A STAND and about 120
ft from a 700 MHz Ascend magnet. Do we have to worry about the vibration
messing up the magnets (we are not worried about experiments currently running)?

Thank you.

Thomas

Thomas K. Weldeghiorghis, Ph.D.
Director, NMR Facility
Department of Chemistry
Louisiana State University
232 Choppin Hall
Baton Rouge, LA 70803
225-578-5346

--
Martha Morton, PhD
Director of Research Instrumentation, University of Nebraska Lincoln
Safety Co-Chair, Department of Chemistry
Nebraska ACS, Chair 2021
ACS, MWRM Chair 2021


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1632): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1632__;!!PvDODwlR4mBZyAb0!SQknFMmuuNOKEUiciPKrnpY1OCAPnO=
ZeJ6b_-QgAyJdOMt8PyfeOGJJMf-L6NOla9WHBw4fW0QtnZq0d$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/108565559=
/7559972__;!!PvDODwlR4mBZyAb0!SQknFMmuuNOKEUiciPKrnpY1OCAPnOZeJ6b_-QgAyJdOM=
t8PyfeOGJJMf-L6NOla9WHBw4fW0bgSrHf3$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Sun Sep 22 2024 - 14:39:32 MST

This archive was generated by hypermail 2.4.0 : Tue Oct 01 2024 - 14:06:56 MST