Dear Shawn,
do you mean users like in "Linux users" (which requires Linux root access level)
or "only" VnmrJ Operators (VJ "vnmr1" level)?
The latter might be solved by copying the contents of /vnmr/adm/users, then
run Tools - Update Locator - Update All
The former is a bit more tricky in terms of getting the same user IDs on all
computers - essentially you'd need to copy the relevant nmr user's section
of /etc/passwd, /etc/shadow and /etc/group, then create the user directories
in /home and run "makeuser XYZuser" as vnmr1 for each of them, possibly
from a short shell script.
Best regards
Bert
> Am 09.09.2024 um 20:47 schrieb Shawn Wagner <shawn.wagner_at_usc.edu>:
>
> Has any attempted to replicate what the VnmrJ Admin program does?
>
> We have 5 different NMR computers and would like to be able to add the same
> user to each instrument computer without have to do it individually, so
> a script that can ssh into each computer and perform the necessary file
> manipulations and permission changes would be quite useful.
>
> Don't want to reinvent the wheel if someone has done this.
>
> Thanks,
> Shawn
>
>
> Shawn Wagner, PhD
> Director of Chemistry Instrumentation Resources
> Laird J. Stabler Memorial Hall
> LJS 160
> 840 Downey Way
> Los Angeles, CA 90089
>
Magnetic resonance equipment, service, training & more
Selling & buying used NMR equipment (all vendors)
Dr. rer. nat. Bert Heise
Owner, Applications Scientist
Spin-Doc Bert Heise
Kiefernweg 13
58239 Schwerte
Germany
Mobile: +49 171 5135315
Email: bert.heise_at_spin-doc.net
Web: http://www.spin-doc.net
Representing:
Stelar Fast Field Cycling NMR Relaxometers (
http://www.stelar.it)
MestreLab NMR Software (
http://www.mestrelab.com)
Barthel HF Amplifiers (
http://www.barthel-hf.de
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1609): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1609__;!!PvDODwlR4mBZyAb0!VZ22VAjhZMB_C9Ihz1R30NnM0fD78a=
IU2FdJH1V74mDr5ffjnN85bB8vjmh2O1_n7N7bmuIKIWCJxt4aczSC2A$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/108360909=
/7559972__;!!PvDODwlR4mBZyAb0!VZ22VAjhZMB_C9Ihz1R30NnM0fD78aIU2FdJH1V74mDr5=
ffjnN85bB8vjmh2O1_n7N7bmuIKIWCJxt7DxxaL4Q$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Tue Sep 10 2024 - 03:16:23 MST