Re: [AMMRL] Tubing for N2 fill exhaust?

From: Steve Clemens via groups.io <steven.clemens=novachem.com_at_groups.io>
Date: Wed, 04 Sep 2024 11:12:38 -0700

The magnet ought to have a pressure release valve on one of the nitrogen turrets
to handle overpressure situations in the magnet.  It may be integrated with
one of the heat exchanger caps.  So I would tend to agree with Perry on this
one.  The pressure safety device shouldn't put people at risk (which is the
case with a rubber hose).


-=-=-=-=-=-=-=-=-=-=-=-