AMMRL: Inova 400 multiple problems

From: Jeff de Ropp <jsderopp_at_ucdavis.edu>
Date: Thu, 14 Sep 2006 16:05:32 -0700

Dear AMMRL:
we have an Inova 400 showing multiple problems. We have described them all
in case in some way they are related.

First, when attempting to start a data acquisition or a mtune we receive
the following error message:
"Rcvr 1: Number of points acquired is # less than np"
where the # is equal to 248 or 524296.
Second, the lock scan does not have typical sinusoidal oscillations and
appears to not update the swept signal fast enough.
Third, occasionally the bottom left side status panel becomes yellow and
reads "inactive" for no apparent reason.

We have tried the standard things: su acqproc, verified connections between
the computer and the NMR console, done a power cycle on the console and
reboot of the computer.

Any ideas?
(I will post a summary if there is interest).
Jeff


Jeffrey S. de Ropp, Ph.D.
NMR Facility
UC Davis
One Shields Avenue
Davis CA 95616
jsderopp_at_ucdavis.edu
530-752-7677 office
530-752-6480 MS1-D Lab
530-752-5682 600 MHz Lab
530-754-8238 Chem NMR Lab
530-752-3516 Fax
NMR Facility Web site URL is:
http://www.nmr.ucdavis.edu/
Received on Thu Sep 14 2006 - 18:08:51 MST

This archive was generated by hypermail 2.4.0 : Sun Jun 11 2023 - 14:55:34 MST