Re: NMR safety: Oxygen sensors in your laboratory

From: David Wilbur <dwilbu01_at_emerald.tufts.edu>
Date: Mon, 25 Sep 2000 16:42:23 -0400

 Oxygen sensors designed for MRI use are made by Enmet (model ISA-40M ).
See http://www.enmet.com/medical.html
They are standard equipment in most MRI facilities. As I recall the
price is about $1500. The sensor cell has a limited lifetime, about 1
year. Replacement cells on the order of $100. They work well if
maintained. As the cell ages, they must be periodically recalibrated,
or the alarm eventually goes off at normal oxygen levels.

Dave