[AMMRL] Tuning issues

From: Eugenio Alvarado via main <ealva=umich.edu_at_groups.io>
Date: Fri, 30 Oct 2026 15:16:34 -0400

Hi all,

Our Varian Vnmrs 500 is intermittently having the problem that the probe
cannot be tuned using autotune or ptcal.  Running protune may take 30
seconds or more to show any reflection data on its window. When it takes
more than 30 seconds the window closes, but sometimes several seconds
later you can hear the tuning motors going on and then it tunes
properly.  Some other times it takes so long that tuning completely
quits and the "su" command is unresponsive for up to a minute, but then
it comes back to normal.  It is like it is very slow getting reflection
data and that puts other processes on hold.  Something similar happens
with ptcal or mtune.  This behavior may continue for several hours and
then work properly for days or weeks.  Testcontrollers show no dropped
packets in the ethernet switch.  Voltages from the power supply are all
correct.

Any ideas will be greatly appreciated.

Eugenio

--
Eugenio Alvarado, Ph.D.
NMR Facility Director
Chemistry Department, room 3500. University of Michigan
930 N. University Ave., Ann Arbor, MI 48109-1055.
Voice: (734)-764-9924


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#3899): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/3899__;!!PvDODwlR4mBZyAb0!WYBzLdQcXuuH-sKuBGryAibcHfrfew=
V-k1MXE_a_Kw_qrve8BvQghSgdw4ajutPZWdfQoJS9QV1wAmzIlL17NxJF13w$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/122017865=
/7559972__;!!PvDODwlR4mBZyAb0!WYBzLdQcXuuH-sKuBGryAibcHfrfewV-k1MXE_a_Kw_qr=
ve8BvQghSgdw4ajutPZWdfQoJS9QV1wAmzIlL17TwJUWY0$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Fri Oct 30 2026 - 12:53:11 MST

This archive was generated by hypermail 2.4.0 : Fri Oct 30 2026 - 14:48:12 MST