Hi Shawn,

One of the usual suspects would be the main power supply - check especially the 5 V supplies in
DC & AC mode of your multimeter (best 5.2 V DC, max= . 20 mV or so AC). Be sure to use a crossover
ethernet cable (usually those grey ones with blue connectors) - while PC network cards are auto-detecting,
the MVME cards aren’t and may get confused.

Best regards 
Bert

-> Sent via mobile <-

Spin-Doc 
Magnetic resonance equipment, service, training & more
Dr. rer. nat. Bert Heise
Owner, applications scientist
Kiefernweg 13
58239 Schwerte
Germany
Mail: bert.heise_at_spin-doc.net
Phone: +49-171-5135315

Representing MestreLab software, Stelar relaxometers & Barthel HF amplifiers



Am 22.10.2026 um 01:03 schrieb Shawn Wagner via groups.io <shawn.wagner=usc.edu_at_groups.io>:

I am trying to revive an NMR (Unity Inova) but I am having a communication issue. After I run setacq there is an immediate connection, but it does not last. If I ping 172.16.0.2, I can get 1 or 2 packets before communication is dropped. Does anyone know why this may happen? Without the connection, OpenVnmrJ just shows inactive.