Hi all,
I'm trying to add HFIP-d2 as a solvent to my 500 MHz Avance NEO with a cryoprobe
(running Topspin 4.5.0 on a Windows 11 PC). I've successfully added it as
a lock solvent but "topshim solvcal" fails with the error message (screenshot attached):
"In experiment: 1D_acquisition_data 3 1 C:/Bruker/TopSpin4.5.0/prog/curdir/NMR_User/topshimData
(Parameter Set=")
unknown exception in drucontr-TRX2 procedure"
I have a ticket going with Bruker for this issue but I'm wondering if anyone has
seen this error message? It could be related to file permissions but I've
ensured that the NMR_User account is a member of the PC Administrators group as
well as a member of the NmrUser and NmrSuperUser groups. I've also tried
explicitly running Topspin as an administrator.
Regular "topshim" completes without errors. Just seems to have a problem
with "solvcal".
Thanks for any help.
Wendy
-=-=-=-=-=-=-=-=-=-=-=-