From: Zhao, Chunqing via groups.io <czhao1=unc.edu_at_groups.io>
Date: Tue, 14 Oct 2025 16:16:53 +0000

Hi all:

The problem was Z0/Z1 board. I replaced it with another board. The system is
working again.

I would like to thank Bert (Spin-Doc); Kereszles Ivan; John Hahn, and Dan
Frank for their very useful advice.

Nelson
UNC ESOP NMR lab

> From: main_at_ammrl.groups.io On Behalf Of Zhao, Chunqing via groups.io
> Sent: Monday, October 6, 2025 2:46 PM
> To: main <main_at_ammrl.groups.io>
> Subject: [AMMRL] find z0 problem on INOVA 400

Hi:

I have a problem on our INOVA 400 which happened over the weekend.

I inserted Agilent Auto Test Sample and clicked Find z0 button from Start/Lock
page. I enabled Lock scan and got a flat lock scan if Lock box was unchecked,
but a sine wave when Lock box was checked. I tried to calibrate Gradient,
generate a new lk gradient map, and calibrate z0 with Auto Test Sample but
didn’t help. Any idea what was wrong?

The system is INOVA 400 with AutoX probe ( We don’t have broad band
probe for this system. Just use AutoX probe to run 1H and 13C related experiments).
OpenVnamrJ 2.1 Revision A is the NMR software which is installed on a centOS
Linux 7.9.2009.

Thanks,

Nelson
UNC ESOP NMR lab



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2925): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2925__;!!PvDODwlR4mBZyAb0!SLDBaP33p4_7WwvfITGJ6gSxQKi-e-=
ZGsCxHCk8dDUzY3XUsB3MZVqwBfZu47aaCKyVCvQA7xBTxERWUqF4g3-1HsQ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/115623804=
/7559972__;!!PvDODwlR4mBZyAb0!SLDBaP33p4_7WwvfITGJ6gSxQKi-e-ZGsCxHCk8dDUzY3=
XUsB3MZVqwBfZu47aaCKyVCvQA7xBTxERWUqF4MQkqHKw$
Group Owner: main+owner_at_ammrl.groups.io
Unsubscribe: https://urldefense.com/v3/__https://ammrl.groups.io/g/main/lea=
ve/12177410/7559972/1125599165/xyzzy__;!!PvDODwlR4mBZyAb0!SLDBaP33p4_7WwvfI=



Received on Tue Oct 14 2025 - 09:42:28 MST

This archive was generated by hypermail 2.4.0 : Sat Nov 01 2025 - 15:23:01 MST