Hi Wendy,
I use a solution of LiCl in D2O. But be aware of the long T1 (30+ seconds).
I often add a trace of CuSO4. That brings the T1 down to ~ 1 - 2 seconds and
turns the solution the color of Mountain Dew.
Clemens
On 10/7/2025 11:00 AM, Wendy Breyer via groups.io wrote:
Hi,
Does anyone have a recommendation for a compound to use to calibrate a 7Li
90-degree pulse?
TIA,
Wendy
--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 bluesky: _at_canklin.bsky.social
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com
==========================
==========================
====================
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.