Hi Jim,
Probably you will hear from Bruker, but I was strongly advised by them against
doing this when we had an issue with an atm probe (air line problem).
Regards,
K. Klika
________________________________________
> From: main_at_ammrl.groups.io on behalf of Baleja, James D via groups.io
> Sent: 03 October 2025 22:39:41
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] Tuning rods pinned?
Hi all, I reset the PICS line, and re-wrote the .bb file (changing the +244
to -99, because that’s what I saw in the 90%/D2O line)
But alas get a similar error message "Can't move to position-99 (current 3)"
Maybe the next step is to remove the ATM unit so I can access the tuning rods
(I don’t think they are accessible without removing the ATM).
First I have to figure out how to remove the ATM unit!
Have a good weekend
Jim
--------------------------
Jim Baleja
Pronouns: he, him, his
Department of Medical Education (MV525S)
Department of Developmental, Molecular, and Chemical Biology (lab:Room MV614, rear)
Tufts University School of Medicine
136 Harrison Avenue
Boston, MA 02111
Tel: 617-636-6872
> From: main_at_ammrl.groups.io On Behalf Of PLscientific
> Sent: Thursday, October 2, 2025 6:51 PM
> To: main_at_ammrl.groups.io
> Subject: [External] Re: [AMMRL] Tuning rods pinned?
Hi Jim,
I would repeat the Ed suggestion. It is indeed a probe-dependent, but 244 looks
too high a number for 1H tuning, so the .bb file was likely overwritten with
the wrong number. In whichever case, after correction of .bb file (if it was
done correctly for the correct NMR probe), you should see a different number
in your ATMA error, like "Can't move to position 21 , current -3" . If it
continues to complain about 244, restart Topspin one more time. Indeed, it's
good to swap PICs because the ATMA probe reinitialization will be done again afterward.
Regarding the .bb file, other solvents may have a correct value, so you may
try to change a solvent simply by starting lock on another solvent.
If a similar problem occurs with the correct .bb file, the ATMA system will
have to be repaired. The simplest would be to check if the motor has not
got stuck in one of the limit positions (a rare reason, but you mentioned 19F).
Over the last three years, > 10 of our customers have asked us to fix that
error-related fault, so it is relatively common for the 2nd generation of ATMA probes...
Regards,
Dr. Pavel Iouchmanov,
--
Repair/upgrade of NMR probes and NMR spectrometers
Electrophoretic NMR and custom NMR solution for electrochemistry
--
+46736220672
P&L Scientific AB
Applied Physical Chemistry, KTH
Teknikringen 36,
100 44 Stockholm
Sweden
pavel_at_plscientific.se<mailto:pavel_at_plscientific.se>
http://www.plscientific.se
-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#2883): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/2883__;!!PvDODwlR4mBZyAb0!VcPImDzfN3_CACyBcFMIfpPF3MfggI=
aWVX-7vSkuykaLDsITJ5qM0PK5CDN4uI-lUvFGNOMhh9NUbBNXJED3J5xM$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/115545023=
/7559972__;!!PvDODwlR4mBZyAb0!VcPImDzfN3_CACyBcFMIfpPF3MfggIaWVX-7vSkuykaLD=
sITJ5qM0PK5CDN4uI-lUvFGNOMhh9NUbBNXJH6HUDqa$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-
Received on Fri Oct 03 2025 - 13:29:42 MST