Re: [AMMRL] Quantum Technology all-metal helium recovery system questions: transport dewar and transfers

From: Xia, Youlin via groups.io <youlin.xia=stjude.org_at_groups.io>
Date: Mon, 21 Oct 2024 16:05:00 +0000

Hello, Margaret,

We installed a QT all-metal system for six magnets (1.1G, 850, 2x800, 2x700)
in April of this year. It has been running well for 6 six months. Both our
mother Dewar and transport Dewar are 250L. It does take several hours to
reduce the pressure in mother Dewar from 6.7 psi to 2 psi, but we transfer
earlier. Normally we start the pre-transfer at home around 8AM, then when t=
he pressure drops to less than 4psi in afternoon we start the transfer. Of
course if you wait for getting 2psi, you may need to transfer in evening.

We installed a computer for the system. Then we can use NoMachine or VNC to
remotely control the system. So we can start the pre-transfer process at home.

We have a scale to get the weight of transport Dewar. The loss of the transport
Dewar is about 2L/day on average . Your Dewar may have bad vacuum or leak.

By the way, another QT all-metal system will be delivered in December. We
like the system. Our recovery rate should be more than 90% since we have not
ordered external Helium for six months.

Best regards,
Youlin
St Jude Children's Research Hospital


________________________________

Email Disclaimer: http://www.stjude.org/emaildisclaimer
Consultation Disclaimer: http://www.stjude.org/consultationdisclaimer


-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1775): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1775__;!!PvDODwlR4mBZyAb0!SXw09TBVQcQmVv5tpx-Nqdtg8pkvNn=
zd9bJyRb8c21LDA9AD083esc2hSX_P_NqOwk1SfB7QwNSZth-TmHrlodH-xIpj$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/109133895=
/7559972__;!!PvDODwlR4mBZyAb0!SXw09TBVQcQmVv5tpx-Nqdtg8pkvNnzd9bJyRb8c21LDA=
9AD083esc2hSX_P_NqOwk1SfB7QwNSZth-TmHrloQAZ-Y6Y$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Mon Oct 21 2024 - 09:16:04 MST

This archive was generated by hypermail 2.4.0 : Fri Nov 01 2024 - 17:08:12 MST