Re: [AMMRL] communication problem with the console

From: Anklin, Clemens via groups.io <clemens.anklin=bruker.com_at_groups.io>
Date: Sun, 20 Oct 2024 21:48:53 +0000

Hi Karel

make sure all the necessary services are running: bootparamd, the Bruker
DHCP server and Tftp

Clemens

On 10/20/2024 12:25, Karel Klika via groups.io wrote:

Dear all,

Our computer and console are not communicating and the postcode on the IPSO
board of our Avance III is showing CO. Anyway to resolve this ?

Thanks in advance,
K. Klika


--
==========================
==========================
====================
Clemens Anklin Ph.D. Vice President
Bruker BioSpin NMR Applications & Training
15 Fortune Drive NEW Phone: 978-313-5NMR(5667)
Billerica MA 01821 twitter: _at_canklin
web: http://www.bruker.com
e-mail: clemens.anklin_at_bruker.com<
==========================
==========================
====================


-=-=-=-=-=-=-=-=-=-=-=-