Re: [AMMRL] AMMRL: Bruker MAS probe rotor ejection problem

From: Struppe, Jochem via groups.io <Jochem.Struppe=bruker.com_at_groups.io>
Date: Tue, 8 Oct 2024 17:08:49 +0000

Weiguo,

In that case if it stops working ,did you try and reset the MAS controller
– i.e. power cycle the unit? I typically advice against such drastic
measures, but it may help finding the cause.

Jochem



> From: main_at_ammrl.groups.io On Behalf Of weiguo.hu via groups.io
> Sent: Tuesday, October 8, 2024 12:03 PM
> To: main_at_ammrl.groups.io
> Subject: Re: [AMMRL] AMMRL: Bruker MAS probe rotor ejection problem


Brian: I checked with a flashlight from the top of the magnet and found that
the stator does flip.

Jochem: I checked the MAS II settings, the masdisp configuration settings and
edhead. There was no setting item that mentions eject. The one thing I do
know is that the eject function sometimes works (it actually is working as
of now, but I am pretty sure that it will stop working soon), which must
mean that the probe settings are fine.

Thanks,
Weiguo



-=-=-=-=-=-=-=-=-=-=-=-
Groups.io Links: You receive all messages sent to this group.
View/Reply Online (#1738): https://urldefense.com/v3/__https://ammrl.groups=
.io/g/main/message/1738__;!!PvDODwlR4mBZyAb0!WoUEegSFXvSIzrjc45AfowVhjZnaLL=
cAEBEqS3hMjz-g4Zbolp7geMybNODEvtXoH_6C7HTRydRGMeXm1Xs7o0sdL8CJ$
Mute This Topic: https://urldefense.com/v3/__https://groups.io/mt/108868672=
/7559972__;!!PvDODwlR4mBZyAb0!WoUEegSFXvSIzrjc45AfowVhjZnaLLcAEBEqS3hMjz-g4=
Zbolp7geMybNODEvtXoH_6C7HTRydRGMeXm1Xs7o8ha_n_6$
Group Owner: main+owner_at_ammrl.groups.io
-=-=-=-=-=-=-=-=-=-=-=-




Received on Tue Oct 08 2024 - 10:09:12 MST

This archive was generated by hypermail 2.4.0 : Fri Nov 01 2024 - 17:08:12 MST