Re: AMMRL: Unexpected shutdown of CNS61-D compressor

From: David H Live <dlive_at_ccrc.uga.edu>
Date: Mon, 30 Oct 2017 17:51:41 +0000

The coils can look clean but there can be a coating of fine dirt or soot on them that adversely affects heat exchange. It is good to periodically apply liquid coil cleaner of the kind used for home air conditioning units and hose the coil off. The water usually comes off looking rather dirty. The stuff is readily available at hardware stores. Also, I think your compressor is a CNA61 not a CNS61.


David Live

Senior Research Scientist
Complex Carbohydrate Research Center

University of Georgia

315 Riverbend Rd.

Athens, GA 30602

706-542-6262


________________________________
From: Thomas Eubanks <thomas.eubanks_at_utrgv.edu>
Sent: Sunday, October 29, 2017 10:49 PM
To: Rainer Haessner; 'ammrl_at_ammrl.org'; Dr. Gerd Gemmecker; Stehle, Ralf; Asami, Sam
Subject: Re: AMMRL: Unexpected shutdown of CNS61-D compressor

I have found Strip connectors where wires are crimped onto tube and pin pairs cause all kinds of problems that seem to be heat related. It is a major fault point in old video arcade games and pinball machines also in more modern centrifuges. Inspect all connections closely both visually and with thermal images with a hand gun. If you use a funnel to focus a heat gun onto small areas it can sometimes expose weak connections. Mouse urine may have caused corrosion. Most of this requires the equipment be powered up...happy hunting!
________________________________
From: Rainer Haessner <rainer.haessner_at_tum.de>
Sent: Saturday, October 28, 2017 4:46:09 AM
To: 'ammrl_at_ammrl.org'; Dr. Gerd Gemmecker; Stehle, Ralf; Asami, Sam
Subject: AMMRL: Unexpected shutdown of CNS61-D compressor


Dear cryoprobe specialists,



we have a very funny malfunction of an about 45 000h old air cooled helium compressor CNS61-D.

We already ordered a new one, nevertheless I am very, very curious, what's going on.



The obervation: as soon as the outside temperature falls below a limit somewhat around 5 and 6 degree centigrade,

the compressor switches off without any warning.



Of course we made some checks. The very first one was cleaning the heat exchanger. There is a sensor outside of

the heat exchanger, which switches the cooling fan inside the housing to lower speed at ... 5 degree centigrade.

If the heat exchanger is not clean, this might result in overpressurized helium gas, which definitely is a shutdown

condition.

Very easy ... but wrong. The heat exchanger is perfect clean, using ice and hot water we triggered the temperature switch to change

the fan speed multiple times without any effect. Even more, to be really sure, we deactivated the temperature sensor

to ensure, the cooling fan is running with full speed all time.



No effect, as soon as the outside temperature falls below 5 ...6 degree centigrade ... see above.



There is a second temperature sensor inside the cabinet. At temperatures below 4 degree centigrade this switches on

6 wires to heat the oil inside the compressor. These wire are driven by 230V (at least in Europe). Now I have to

append, that we had a mouse living inside the cabinet. This mouse caused some minor destruction work at these wires.

At some places the insulation shows teeth marks from the mouse. It is very difficult to see, whether there was a significant

damage or not. But anyway, tentatively we deactivated this sensor as well.



No effect, deeper outside temperature, sudden breakdown of the compressor.



There is a low pressure sensor for the return gas. Deactivated ... no effect.



I am out of ideas, replacement compressor will arrive soon, nevertheless that's a real challenge. There *has to be* a

deterministic reason.



Looking for any idea ...



Greetings



Rainer