Re: AMMRL: VnmrJ error "Unable to find Acode Set"

From: Allan Kershaw <adk_at_nmrnet.usc.edu>
Date: Tue, 20 Oct 2015 08:25:49 -0700

Alvaro,

I have had this error and it has been discussed in the spinsights Help.
Below is the response from Dan Iverson on this problem:

The most common cause for this is a mis-configured system. The config
command lets you specify the hardware that is in the system. It must
match the actual hardware. For example, if you have 3 RF channels and
config only specifies 2, you will get this error. From a teminal window,
enter

    rlogin master1

Then run the acquisition from VJ. After you get the error, in the
terminal window there will be a message similar to

       'master1': Exception from: -> 'grad1' <-, Type: 15, ErrorCode: 1204

Where grad1 could be rf2, pfg1, ddr2, etc. That will tell you which
controller is not getting its acodes. Make sure config is set correctly.
To exit from the master1 login, enter

       logout

Dan


Hopefully this will help.

Allan Kershaw
Director of Advanced Instrumentation
Department of Chemistry
University of Southern California
840 Downey Way LJS 159
Los Angeles CA 90089
213 740-6376

On 10/19/2015 11:28 AM, Alvaro Herrera wrote:
>
>
>
>
> Hello Everyone,
>
>
> Perhaps someone out there has faced the same problem we are having. We
> are running Vnmrj 3.2A (with all patches) on our 500 MHz instrument.
> When running anything different from a proton 1D I am getting the
> following error "Unable to find Acode Set", it comes after finishing
> acquisition, and there is no data after it. I would appreciate any ideas.
>
>
> Thanks
>
>
>
> Alvaro Herrera
>
> NMR Facility Manager
>
> Biochemistry and Molecular Biophysics
>
> Kansas State University
>
>
>
>
Received on Tue Oct 20 2015 - 05:25:49 MST